510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2533-2016Class II

Synovis VASCU-GUARD Peripheral Vascular Patch in the following sizes and product codes: 1 cm x 6 cm, product...

Baxter Healthcare Corporation / initiated 2016-06-24 / Terminated

Baxter Healthcare Corp. began a recall of Synovis VASCU-GUARD Peripheral Vascular Patch in the following sizes and product codes: 1 cm x 6 cm, product code VG0106N, 0.8 cm x 8 cm, product code VG0108N... on . FDA classified it as Class II and gives the reason as "Baxter is issuing a safety alert in response to postmarketing reports received for the VASC U-GUARD Peripheral Vascular Patch. Baxter has received reports for intraoperative or postoperative bleeding episodes, which...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2533-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Synovis VASCU-GUARD Peripheral Vascular Patch in the following sizes and product codes: 1 cm x 6 cm, product code VG0106N, 0.8 cm x 8 cm, product code VG0108N, 1 cm x 10 cm, product code VG0110N, and 2 cm x 9 cm, product code VG0209N. VASC U-GUARD is prepared from bovine pericardium which is cross-linked with glutaraldehyde. Sterilized using aseptic processing techniques. The pericardium is procured from cattle originating in the United States. VASC-Guard is chemically sterilized using ethanol and propylene oxide. UPDATE 9/1/2016 Retrieval of devices only includes Synovis VASCU-GUARD Peripheral Vascular Patch in the following sizes and product codes: 1 cm x 6 cm, product code VG0106N, 0.8 cm x 8 cm, product code VG0108N.
Product code
DXZ Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene
Reason for recall
Baxter is issuing a safety alert in response to postmarketing reports received for the VASC U-GUARD Peripheral Vascular Patch. Baxter has received reports for intraoperative or postoperative bleeding episodes, which required additional clinical intervention. One report involved a case with a fatal outcome. UPDATE 9/1/2016 Baxter issued a notice retrieving 6 lots of VASCU-GUARD due to the increased postmarketing reports related to intraoperative and postoperative bleeding.
FDA root cause
Under Investigation by firm
Action
Baxter sent a Safety Alert to customers via USPS overnight delivery beginning 6/24/2016. The letter told of the reported adverse events, asked for product that product immediately be discontinued from use, to segregate/quarantine all affected lots, and to return the enclosed customer reply form. General questions regarding the Safety Alert should be directed to Baxter Product Surveillance at 1-800-437-5176, between 8:00 am - 5:00 pm Central Time Monday thru Friday. Clinical inquiries can be directed to Baxter's Medical information at 1-800-933-0303, Monday thru Friday 8:00 am - 5:00 pm central time. UPDATE 9/1/2016 Baxter issued an Urgent Product Recall letter beginning 9/1/2016. This letter explained the issue and identified affected product that was to be removed from the consignee and returned to Baxter. Customer is to contact Baxter Healthcare Center for Service to arrange for return and credit. This can be done at 888-229-0001 Monday thru Friday 7:00 am - 6:00 pm Central Time. General questions regarding the recall can be addressed to Baxter Product Surveillance at 800-437-5179 Monday thru Friday, 8:00 am - 5:00 pm Central Time.
Quantity in commerce
32,763 eaches UPDATE 9/1/2016 2,042 eaches
Distribution
Nationwide including Puerto Rico, and outside the US to include: Australia, Austria, Finland, Germany, Greece, Ireland, Italy, the Netherlands, Norway, Poland, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, United Kingdom.
Product codes and lots
Product Code VG0106N, Lot #: SP16A04-1108933, SP16A11-1111696, SP16A12-1111933, SP16A12-1111938, SP16A12-1111942, SP16A12-1111947, SP16A27-1116312, SP16A29-1116884, SP16B04-1118465, SP16B08-1119650, SP16B12-1121361, SP16B22-1123979, SP16B24-1124658, SP16C01-1126439, SP16C01-1126508, SP16C07-1128790, SP16C08-1129067, SP16C09-1129905, SP16C10-1130129, SP16C10-1130134, SP16C23-1134739, SP16C28-1136324, SP16C28-1136332, SP16D06-1139370, SP16D13-1141301, SP16D19-1142941, SP16D20-1143649, SP16D20-1143655, SP16D21-1143852, SP16D28-1146476, SP16E04-1148533, SP16E04-1148534, SP16E19-1152299, SP16E24-1153675, SP16F01-1155425, SP16F01-1155678, SP16F01-1155682, SP16F01-1155685.   Product Code VG0108N, Lot #: SP16A04-1108941, SP16A04-1108956, SP16A04-1108974, SP16A06-1110391, SP16A12-1111872, SP16A12-1111952, SP16A14-1112703, SP16A14-1112720, SP16A18-1113372, SP16A19-1114051, SP16A19-1114056, SP16A21-1114800, SP16A21-1114802, SP16A21-1114805, SP16B04-1118471, SP16B04-1118475, SP16B05-1119212, SP16B05-1119217, SP16B08-1119654, SP16B08-1119720, SP16B12-1121365, SP16B12-1121366, SP16B19-1123365, SP16B22-1123983, SP16B25-1125254, SP16B25-1125260, SP16B25-1125268, SP16C01-1126459, SP16C07-1128755, SP16C08-1129301, SP16C10-1130184, SP16C10-1130191, SP16C10-1130200, SP16C10-1130208, SP16C10-1130219, SP16C10-1130227, SP16C16-1132096, SP16C18-1133115, SP16C18-1133138, SP16C18-1133148, SP16C18-1133152, SP16C18-1133155, SP16C18-1133157, SP16C21-1133807, SP16C22-1134481, SP16C24-1135156, SP16C28-1136307, SP16C28-1136436, SP16D01-1137764, SP16D04-1138414, SP16D04-1138417, SP16D12-1140707, SP16D12-1140708, SP16D19-1142958, SP16D19-1142967, SP16D21-1143872, SP16D21-1143878, SP16D28-1146462, SP16D28-1146466, SP16D28-1146471, SP16E04-1148520, SP16E04-1148522, SP16E04-1148549, SP16E05-1148667, SP16E11-1150104, SP16E11-1150110, SP16E11-1150115, SP16E11-1150174, SP16E19-1152326, SP16E19-1152330, SP16E19-1152335, SP16E24-1153278, SP16E24-1153280, SP16E26-1154142, SP16E26-1154146, SP16E27-1154532, SP16E27-1154534, SP16F01-1155407, SP16F07-1157497, SP16F07-1157505, SP16F07-1157513.   Product Code VG0110N, Lot #: SP16A12-1111926, SP16A12-1111931, SP16A20-1114346, SP16B22-1123992, SP16C08-1129312, SP16C08-1129329, SP16C09-1129911, SP16C25-1135841, SP16D11-1140274, SP16D11-1140284, SP16D20-1143641, SP16D20-1143643, SP16D25-1144838, SP16E11-1150202, SP16F07-1157244, SP16F07-1157253, SP16F07-1157260.   Product Code VG0209N, Lot #: SP16A04-1108918, SP16A12-1112283, SP16A12-1112284, SP16A20-1114352, SP16A21-1114790, SP16B15-1121467, SP16B22-1124000, SP16B25-1125139, SP16C08-1129339, SP16C08-1129346, SP16C14-1131192, SP16C16-1132144, SP16D01-1137778, SP16D11-1140290, SP16D18-1142652, SP16D20-1143633, SP16D20-1143638, SP16D28-1146487, SP16D28-1146606, SP16E04-1148530, SP16E11-1150205, SP16E19-1152309, SP16E27-1154668, SP16E27-1154693, SP16F08-1157871.   UPDATE 9/1/2016 the following lots are being retrieved for return to Baxter: Product Code VG0106N: Lot #: SP16C09-1129905. Product Code VG0108N: Lot #: SP16B08-1119720, SP16C18-1133152, SP16D04-1138417, SP16D28-1146462, SP16E04-1148522.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2533-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2533-2016Class IITerminatedVoluntary: Firm initiated