510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2238-2016Class II

2.3MM Tapered Router, Product Number 5407-FA2-023 A router is a cutting accessory used in the following...

Stryker Corp. / initiated 2016-06-24 / Terminated / root cause: manufacturing process

Stryker Instruments Div. of Stryker Corporation began a recall of 2.3MM Tapered Router, Product Number 5407-FA2-023 A router is a cutting accessory used in the following medical applications: Neuro; Spine; Ear, Nose, and... on . FDA classified it as Class II and gives the reason as "A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be more susceptible to breakages. Risk to Health: Injury to critical soft tissue requiring surgical...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2238-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
2.3MM Tapered Router, Product Number 5407-FA2-023 A router is a cutting accessory used in the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/Otology /Neurotology/ Otorhinolaryngology; Craniofacial (bones of the skull and supraorbital region); and Sternotomy. They are intended to be used with the Stryker CORE" system.
Product code
HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
Reason for recall
A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be more susceptible to breakages. Risk to Health: Injury to critical soft tissue requiring surgical intervention may occur.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Stryker sent an Urgent Medical Device Recall Notification letter dated June 24, 2016, to the affected customers via courier, and to Sales Representatives and international affiliates via e-mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments at 269-389-2921.
Quantity in commerce
3,163
Distribution
Worldwide Distribution - US (Nationwide) Netherlands, Canada, Japan, Switzerland, and United Kingdom
Product codes and lots
Lot numbers: 15077027, 15077067, 15077097, 15078087, 15105017, 15108017, 15108047, 15108077, 15108087, 15108097, 15146017, 15146027, 15146037, 15146047

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2238-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2238-2016Class IITerminatedVoluntary: Firm initiated