Broselow/Hinkle, PEDIATRIC EMERGENCY SYSTEM, PINK/RED Oxygen Delivery Module 2/PACK, Product Code 7700RAW2
Carefusion 2200 Inc began a recall of Broselow/Hinkle, PEDIATRIC EMERGENCY SYSTEM, PINK/RED Oxygen Delivery Module 2/PACK, Product Code 7700RAW2 on . FDA classified it as Class II and gives the reason as "The Oral Airway pouch within the PINK/RED Oxygen Delivery Module, 7700RAW, was incorrectly packaged with the 8 cm Berman Oral Airway. The pouch contained within the module is correctly labeled as 5 cm but actually...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2846-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Broselow/Hinkle, PEDIATRIC EMERGENCY SYSTEM, PINK/RED Oxygen Delivery Module 2/PACK, Product Code 7700RAW2
- Product code
- CAE Airway, Oropharyngeal, Anesthesiology
- Reason for recall
- The Oral Airway pouch within the PINK/RED Oxygen Delivery Module, 7700RAW, was incorrectly packaged with the 8 cm Berman Oral Airway. The pouch contained within the module is correctly labeled as 5 cm but actually contains an 8 cm Berman Oral Airway component. Therefore, the component within the pouch is incorrect. The PINK/RED Oxygen Delivery Module, 7700RAW, is intended to be used within the PINK/RED Pouch of the BROSELOW pediatric safety system. Three (3) lots (lot# 914873, lot# 914876 and lot #914878) of the PINK/RED Oxygen Delivery Module, 7700RAW, are affected by this voluntary Field Safety Corrective Action.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Recall notices were sent 6/29/2016.
- Quantity in commerce
- 544 (2/PACKS)
- Distribution
- Illinois
- Product codes and lots
- Lot numbers 914873, 914876, and 914878 of the individual modules (7700RAW), which are contained in Lot number 914867 of the 2/PACK (7700RAW2). Expiration Date JAN 2021.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2846-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2847-2016 | Class II | Terminated | Voluntary: Firm initiated |