Z-2327-2016Class II
Horizon 24 Large Titanium Clips with Adhesive Backing, Ref 004204, Rx Only, Sterile. The metal ligating clips...
Teleflex Medical / initiated 2013-08-08 / Terminated / root cause: labeling and UDI
Teleflex Medical began a recall of Horizon 24 Large Titanium Clips with Adhesive Backing, Ref 004204, Rx Only, Sterile. The metal ligating clips are intended for use in procedures involving... on . FDA classified it as Class II and gives the reason as "The manufacturing and expiration dates are not printed on the blister pack label". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2327-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Horizon 24 Large Titanium Clips with Adhesive Backing, Ref 004204, Rx Only, Sterile. The metal ligating clips are intended for use in procedures involving vessels or anatomic structures for which the surgeon determines ligating clips are the best choice. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable
- Product code
- FZP Clip, Implantable
- Reason for recall
- The manufacturing and expiration dates are not printed on the blister pack label.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Teleflex sent an Urgent Medical Device Recall Notification dated August 8, 2013, to all affected customers. The letter requested that they quarantine the product and to return it to Teleflex. An acknowledgment form was included with the letter which also was to be returned to Teleflex. A separate letter was sent to distributors requesting they conduct a sub-recall. If you have any further questions, please call (610) 378-0131.
- Quantity in commerce
- 8,220 ea.
- Distribution
- USA Distribution to the states of :AL, AZ, CA, CT, FL, IL, KY, MD, MI, MN, NM, NY, NC, OR, PA, SC, TX, VA, WA, WI, and Internationally to Canada
- Product codes and lots
- Lot/Serial Number 01J1200435
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2327-2016 | Class II | Terminated | Voluntary: Firm initiated |