Z-2670-2016Class II
UNIQ 1.0: UNIQ FD10, 722026; UNIQ FD10/10, 722027; UNIQ FD20, 722028; UNIQ FD20/10 biplane, 722029; UNIQ FD10...
Philips North America LLC / initiated 2016-07-07 / Terminated
Philips Electronics North America Corporation began a recall of UNIQ 1.0: UNIQ FD10, 722026; UNIQ FD10/10, 722027; UNIQ FD20, 722028; UNIQ FD20/10 biplane, 722029; UNIQ FD10 OR Table, 722033; UNIQ FD10/10 OR Table, 722034... on . FDA classified it as Class II and gives the reason as "Increase in the failure rate of certain low-voltage DC power supplies (DCPS)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2670-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- UNIQ 1.0: UNIQ FD10, 722026; UNIQ FD10/10, 722027; UNIQ FD20, 722028; UNIQ FD20/10 biplane, 722029; UNIQ FD10 OR Table, 722033; UNIQ FD10/10 OR Table, 722034; UNIQ FD20 OR Table, 722035; UNIQ FD20/20, 722038; UNIQ FD20/20 biplane OR Table, 722039; UNIQ FD20/15, 722058; UNIQ FD20/15 OR Table 722059
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Increase in the failure rate of certain low-voltage DC power supplies (DCPS)
- FDA root cause
- Component change control
- Action
- Philips sent an Urgent - Field Safety Notice dated June 20, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Field Service Engineers will replace the affected DCPS free of charge. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377. For further questions regarding this recall please call (978) 659-3000.
- Quantity in commerce
- 1536 devices (includes both Allura and UNIQ)
- Distribution
- Nationwide, Canada, Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Bulgaria CHILE China Colombia Croatia Cuba Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Finland France French Polynesia Georgia Germany Greece Honduras Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Lebanon Lithuania Macedonia Malaysia Malta Mexico Morocco Myanmar Nepal Netherlands New Caledonia Norway Oman Pakistan Palestine, State of PERU Philippines Poland Portugal Qatar R¿union Romania RUSSIAN FEDERAT Russian Federation Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Taiwan Tanzania Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Viet Nam
- Product codes and lots
- UNIQ with release R1.0
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2669-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2670-2016 | Class II | Terminated | Voluntary: Firm initiated |