Philips DuraDiagnost stationary X-ray system
Philips Electronics North America Corporation began a recall of Philips DuraDiagnost stationary X-ray system on . FDA classified it as Class II and gives the reason as "Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2682-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips DuraDiagnost stationary X-ray system
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable Tabletop is missing the required labels [21 CFR 1010.2., 21 CFR 1010.3(a), and 21 CFR 1020.30(3)].
- Root cause tag
- labeling and UDI
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Philips plans to bring the products into compliance with the Federal standard at no cost to customers by implementing a field correction. 1. You will contact customers and initiate the actions to correct the failure to comply issues. 2. You will attach the warning label to maintain an audible volume, and issue a supplement to the user manual that instructs the user how to adjust the volume for the local noise environment and include a warning that the user shall maintain an audible volume. 3. You will attach the identification label to the affected Height-Adjustable Tabletop. 4. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. If you need any further information or support concerning this, please contact your local Philips representative or Philips Customer Service: 1-800-722-9377.
- Quantity in commerce
- 15
- Distribution
- US Distribution to the states of : ME, NE, OR, CA, TX, NC, WA and OH.
- Product codes and lots
- SN: 140054 140059 150013 150014 150039 150041 150042 150052 150053 150055 150056 150059 150061 150091 150093
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2682-2016 | Class II | Terminated | FDA Mandated |