Avenir Mller Stem 4 Standard Avenir Mller Stem 6 Lateral Hip joint metal/ceramic/polymer semi-constrained...
Zimmer Gmbh began a recall of Avenir Mller Stem 4 Standard Avenir Mller Stem 6 Lateral Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Intended... on . FDA classified it as Class II and gives the reason as "Product mix-up. The Avenir Mller Stem 6 lateral uncemented might be placed in the packaging of the Avenir Mller Stem 4 standard uncemented and vice versa. Potential for delay in surgery, and/or potential for use of...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2292-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Avenir Mller Stem 4 Standard Avenir Mller Stem 6 Lateral Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Intended Use: " Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases. " Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, hemiarthroplasty or total hip replacement (THR). " Acute traumatic fracture of the femoral head or neck " Avascular necrosis of the femoral head. " Avenir Moller Stems are for cementless use only.
- Product code
- LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
- Reason for recall
- Product mix-up. The Avenir Mller Stem 6 lateral uncemented might be placed in the packaging of the Avenir Mller Stem 4 standard uncemented and vice versa. Potential for delay in surgery, and/or potential for use of incorrect implant which may in turn lead dismetry in patient and possibly the need for revision.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall letter dated June 30, 2016 to all affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were instructed to review the notification immediately and ensure affected personnel are aware of the contents and assist Zimmer Biomet sales representative with the quarantine of any affected device. Your Zimmer Biomet sales representative will remove any affected device. Customers were also instructed to complete the Certification of Acknowledgement form and return to [email protected]. Customers with further questions or concerns were instructed to contact the customer call center at 1-800-348-2759 between 8:00 am and 8:00 pm EST. For questions regarding this recall call 800-348-2759.
- Quantity in commerce
- 2
- Distribution
- Nationwide Distribution to OH and NM
- Product codes and lots
- Item Name: Avenir M¿ller Stem 4 Standard Item Number: 01.06010.004 Lot Number: 4022860 Item Name: Avenir M¿ller Stem 6 Lateral Item Number: 01.06010.106 Lot Number: 4023094
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2292-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.