510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2537-2016Class I

Roadrunner Uniglide Hydrophilic Wire Guide, Catalog Numbers and GPN: HPWA-18-260, G30489; HPWA-18-320...

Cook Incorporated / initiated 2016-06-24 / Terminated / root cause: manufacturing process

Cook Inc. began a recall of Roadrunner Uniglide Hydrophilic Wire Guide, Catalog Numbers and GPN: HPWA-18-260, G30489; HPWA-18-320, G30490; HPWA-25-260, G30491; HPWA-35-260, G30492... on . FDA classified it as Class I and gives the reason as "Potential coating contamination with glass particles". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2537-2016
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Roadrunner Uniglide Hydrophilic Wire Guide, Catalog Numbers and GPN: HPWA-18-260, G30489; HPWA-18-320, G30490; HPWA-25-260, G30491; HPWA-35-260, G30492; HPWAS-35-260, G30506; HPW-35-150, G56149; HPW-35-180, G56150; HPWS-35-150, G56152; HPWS-35-180, G56153; HPWA-35-80, G56172; HPWA-35-150, G56173; HPWA-35-180, G56174; HPWA-18-180, G56160; HPWAS-35-150, G56176; HPWAS-35-180, G56177. Used with vascular catheter. Roadrunner UniGlide Hydrophilic Wire Guide is constructed from a steerable, metallic core with a polymer coating. A hydrophilic coating is applied over the radiopaque polymer jacket. Sterile- Individual product is packaged in a Tyvek-film sterilizable outer pouch and boxed in a quantity of five.
Product code
DQX Wire, Guide, Catheter
Reason for recall
Potential coating contamination with glass particles.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Cook Inc. initiated a voluntary recall of the Roadrunner UniGlide Hydrophilic Wire Guides on June 24, 2016 via courier due to the potential for the device to be contaminated with glass particles. The firm issued a press release dated August 12, 2016. Customers were asked to take the following actions: 1. Examine inventory immediately to determine if you have affected products and quarantine affected products. 2. Return affected products to Stericycle Expert Solutions (a third-party recall administration service provider). Use the enclosed label and a copy of the Acknowledgement and Receipt Form to receive a product credit. 3. Even if you do not have affected products on hand, you must still complete the Acknowledgement and Receipt Form and fax it to 866.796.4780 or email it to [email protected]. 4. Report adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. Monday through Friday between 7:30 a.m. and 5:00 p.m. (Eastern Time). Or by email at [email protected]. Should you have any medical questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. For information regarding the recall, please contact Stericycle Expert Solutions at 866.912.9552. Cook is providing notification to distributors and informing them that this notice must be shared with appropriate personnel, including down to the user level, within your organization or to any organization where the potentially affected devices have been transferred.
Quantity in commerce
8,750 devices were distributed in commerce
Distribution
Worldwide distribution. US (nationwide), Australia, Austria, Belgium, Canada, China, Czech Republic, France, Germany, Hong Kong, Hungary, Italy, Jordan, Korea, Spain, South Africa, Taiwan, Turkey, and United Kingdom
Product codes and lots
Roadrunner Uniglide Hydrophilic Wire Guide, Catalog Numbers and GPN: HPWA-18-260, G30489; HPWA-18-320, G30490; HPWA-25-260, G30491; HPWA-35-260, G30492; HPWAS-35-260, G30506; HPW-35-150, G56149; HPW-35-180, G56150; HPWS-35-150, G56152; HPWS-35-180, G56153; HPWA-35-80, G56172; HPWA-35-150, G56173; HPWA-35-180, G56174; HPWA-18-180, G56160; HPWAS-35-150, G56176; HPWAS-35-180, G56177.   Lot numbers 6911522, 6911523, 6911524, 6911526, 6911527, 6911528, 6911529, 6911530, 6911533, 6911535, 6911545, 6911549, 6911551, 6911552, 6919676, 6919678, 6919680, 6935843, 6935844, 6935846, 6935848, 6935849, 6935850, 6935851, 6935852, 6935853, 6935855, 6948854, 6948855, 6948856, 6948857, 6948858, 6948860, 6948861, 6948862, 6955228, 6955229, 6955230, 6955231, 6955232, 6955233, 6955234, 6955235, 6955236, 6955237, 6961752, 6961753, 6961754, 6961755, 6961756, 6961757, 6961758, 6961759, 6961760, 6961761, 6965235, 6965236, 6965237, 6965238, 6965240, 6965469, 6965625, 6965626, 6968777, 6968779, 6975365, 6975367, 6975368, 6981022, 6981026, 6981346, 6981347, 6981349, 6981351, 6986896, 6986897, 6986898, 6988487, 6988488, 6992197, 6992198, 6995077, 6995078, 6995080, 6995082, 6995083, 6997860, 6999759, 6999760, 7000724, 7000725, 7000726, 7000727, 7000730, 7000732, 7000733, 7000735, 7012526, 7012527, 7012528, 7012529, 7012530, 7012840, 7015853, 7015854, 7015855, 7015856, 7015857, 7015897, 7015898, 7015899, 7016252, 7016253, 7019017, 7019018, 7019019, 7019020, 7019022, 7019025, 7019027, 7019491, 7019492, 7019493, 7019494, 7021222, 7021223, 7021224, 7021225, 7021226, 7021227, 7022591, 6948855X, 6955234X, NS6911520, NS6935845, NS6935847, NS6935854, NS6948859, NS6965239, NS6968778, NS6975369, NS6981025, NS6995079, NS6999757, NS6999758, NS6999761, NS7000728, NS7012841, NS7019023, NS7019024, and NS7019495.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2537-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2537-2016Class ITerminatedVoluntary: Firm initiated