Z-0296-2017Class II

Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOPATH ETS-Flex Endoscopic...

Ethicon Endo-Surgery Inc / initiated 2016-07-15 / Terminated / root cause: design

Ethicon Endo-Surgery Inc began a recall of Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general... on . FDA classified it as Class II and gives the reason as "Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0296-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon Endo-Surgery Inc
Product
Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses
Product code
GDW Staple, Implantable
Reason for recall
Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury.
Root cause tag
design
FDA root cause
Device Design
Action
Ethicon sent an Urgent Medical Device Recall letter dated July 15, 2016, to all affected consignees via UPS next day mail. The letter identified the product, the problem, and the action to be taken by the customer. Consignees were instructed to examine their inventory immediately and quarantine any affected product and return to ENDOPATH. If product subject to the recall was further distributed consignees were instructed to notify the other facility and arrange for return of the affected product. Customer were instructed to complete the Business Reply Form confirming receipt of this letter and fax to Stericycle at 1-888-228-5719 or email to [email protected]. Customers with questions were instructed to contact the Ethicon Customer Support Center at 1-877-384-4266. For questions regarding this recall call 513-337-3582.
Quantity in commerce
213
Distribution
Worldwide Distribution - US including AL, AR, AZ, CA, CO,CT, DC, FL,GA, IA,IL,KS,KY,LA,MA,MD, ME, MI,MN,MO, MS, MT,NC, ND, NH,NJ, NM, NV, NY, OH, RI, TN, TX, VA, VT, WA, WI and Internationally to Brazil and Canada.
Product codes and lots
N91L2N, N91L2P, N91N1J, N91N1K, N91L0T, N91L0U, N91N29,  N91N2G, N91R8X, N91R8Y.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0296-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0295-2017Class IITerminatedVoluntary: Firm initiated
Z-0296-2017Class IITerminatedVoluntary: Firm initiated