Z-2349-2016Class II

STAPLER 30, IS4000, model number 470430; Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist...

Intuitive Surgical, Inc. / initiated 2016-07-14 / Terminated / root cause: software

Intuitive Surgical, Inc. began a recall of STAPLER 30, IS4000, model number 470430; Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist Stapler 30 Instrument and Stapler 30 Reloads are... on . FDA classified it as Class II and gives the reason as "An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System logs found an increased trend". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2349-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
STAPLER 30, IS4000, model number 470430; Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist Stapler 30 Instrument and Stapler 30 Reloads are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device is indicated for adult and pediatric use. The device can be used with staple line or tissue buttressing material.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System logs found an increased trend
Root cause tag
software
FDA root cause
Software design
Action
Field Safety Notifications - Urgent Medical Device Corrections and Removal letters were sent on July 15, 2016 (dated July 14, 2016 advised users of the Stapler Firing Failed message and actions to take should the message be received.
Quantity in commerce
2065 total
Distribution
US, including Puerto Rico and Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Monaco, Netherlands, Norway, Portugal, , Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Venezuela.
Product codes and lots
All

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2349-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2348-2016Class IITerminatedVoluntary: Firm initiated
Z-2349-2016Class IITerminatedVoluntary: Firm initiated
Z-2350-2016Class IITerminatedVoluntary: Firm initiated