Z-2792-2016Class II

Femoral Nail, A/R T2 Femur 011x280 mm T2 Femoral Nailing System is one of the first femoral nailing systems...

Howmedica Osteonics Corp / initiated 2016-06-30 / Terminated / root cause: labeling and UDI

Stryker Howmedica Osteonics Corp. began a recall of Femoral Nail, A/R T2 Femur 011x280 mm T2 Femoral Nailing System is one of the first femoral nailing systems to offer an option for either an antegrade or a... on . FDA classified it as Class II and gives the reason as "Stryker discovered that a Tibial Nail and femoral Nail were incorrectly labeled and distributed to the field". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2792-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Femoral Nail, A/R T2 Femur 011x280 mm T2 Femoral Nailing System is one of the first femoral nailing systems to offer an option for either an antegrade or a retrograde approach to repair fractures of the femur. (1) Open and closed femoral fractures, (2) Pseudoarthrosis and correction, (3) osteotomy, (4) Pathologic fractures, impending pathologic fractures, and tumor resections, (5) Supracondylar fractures, including those with intra-articular extension, (6) Ipsilateral femur fractures, (7) Fractures proximal to a total knee arthroplasty, (8) Fractures distal to a hip joint and (9) Non-unions and malunions
Product code
HSB Rod, Fixation, Intramedullary And Accessories
Reason for recall
Stryker discovered that a Tibial Nail and femoral Nail were incorrectly labeled and distributed to the field.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
The firm, Stryker, sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 30, 2016 by e-mail to its customer requesting that they quarantine affected devices and sent the letter via UPS (with return receipt) on July 1, 2016. The letter described the product, problem and actions to be taken. The customer was instructed to inform users of the Medical Device Removal and forward the notice to all who need to be aware; inspect your inventory and return affected products to Stryker Orthopaedics/PFA Product Returns, Attn: Distribution Inventory Team; 325 Corporate Drive; Dock M-East , Mahwah, NJ 07431 REF: PFA #2016-094; complete and return the Business Reply Form via email to [email protected] within 5 days; and keep a copy of the completed and executed Business Reply Form for your records. If you have any questions, feel free to contact the Mgr., Divisional Regulatory Compliance, Marketing Quality at (201) 831-6693.
Quantity in commerce
1 unit in total
Distribution
US distribution to state of: CA.
Product codes and lots
Catalog number 18251032S and Lot Code K288335

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2792-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2791-2016Class IITerminatedVoluntary: Firm initiated
Z-2792-2016Class IITerminatedVoluntary: Firm initiated