510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2531-2016Class II

iConnect Access used with Merge PACS software provides medical specialists with access to diagnostic quality...

Merge Healthcare Incorporated / initiated 2016-01-30 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of iConnect Access used with Merge PACS software provides medical specialists with access to diagnostic quality images, reports, and various types of patient data... on . FDA classified it as Class II and gives the reason as "Studies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2531-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
iConnect Access used with Merge PACS software provides medical specialists with access to diagnostic quality images, reports, and various types of patient data over conventional TCP/IP (e.g., internet) networks.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Studies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after the images were ingested in PACS.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued a letter dated 1/28/2016 via e-mail on 1/31/2016 or via certified mail if they did not have an e-mail address for the customer.
Quantity in commerce
146 sites potentially have the affected versions for both products
Distribution
Distribution was made to medical facilities nationwide and to one foreign medical facility in New Zealand. There was no military or government distribution.
Product codes and lots
iConnect Access V4.0 or earlier with any version of Merge PACS and iConnect Access V4.1 with Merge PACS V6.5.4 or earlier

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2531-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2531-2016Class IITerminatedVoluntary: Firm initiated
Z-2532-2016Class IITerminatedVoluntary: Firm initiated