Z-2396-2016Class II

Orthopedic manual surgical instrument Usage: Exeter Rasp Handle is a reusable instrument which is part of the...

Howmedica Osteonics Corp / initiated 2016-07-01 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of Orthopedic manual surgical instrument Usage: Exeter Rasp Handle is a reusable instrument which is part of the Exeter femoral hip system. Exeter femoral rasps... on . FDA classified it as Class II and gives the reason as "Malfunctioning of the Exeter Rasp/trial introducer/extractor Handle". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2396-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Orthopedic manual surgical instrument Usage: Exeter Rasp Handle is a reusable instrument which is part of the Exeter femoral hip system. Exeter femoral rasps are assembled with the rasp handle and locked in place by closing down the lever on the rasp handle. The rasp handle is then used to guide the rasp into the patients femur and is impacted using a mallet in order to remove bone and further prepare the femur.
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
Malfunctioning of the Exeter Rasp/trial introducer/extractor Handle.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker notified Branches/Agencies of the Urgent Medical Device Recall by e-mail on July 1, 2016 and an "Urgent Medical Device Recall Letter and Business Reply Form" was sent to branches, agents and hospital risk management via UPS (with return receipt) on July 5, 2016. The letter identified then affected product, problem and actions to be taken. For questions call (201) 831-6693.
Quantity in commerce
241 units
Distribution
Worldwide Distribution - US Nationwide
Product codes and lots
Catalog No. 0930-9-003

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2396-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2396-2016Class IITerminatedVoluntary: Firm initiated