Radiation Therapy Treatment Planning System, Model 5.0
RAYSEARCH LABORATORIES AB began a recall of Radiation Therapy Treatment Planning System, Model 5.0 on . FDA classified it as Class II and gives the reason as "For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0079-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- Radiation Therapy Treatment Planning System, Model 5.0
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On May 19th, 2016, RaySearch Laboratories distributed Field Safety Notice, Medical Device Correction notices and reply forms to their customers via email. Actions to be taken by the customer include: 1) Inspect all report templates that are intended to be used for plans with multiple beam sets. 2) Make sure that POI Dose statistics (Plan) is included immediately before ROI Dose statistics [Plan dose] in the Report content area. 3) Please educate planning staff and all users about this workaround. 4) Inspect your product and identify all installed units with the above software version number. Finally, complete and return the reply form via email to: [email protected] or by fax to: 888-501-7195 to confirm that you have read and understand the information. Customers with questions can contact David Hedfors via email: [email protected]
- Quantity in commerce
- 491 units (187 domestically & 304 internationally)
- Distribution
- AZ, CA, FL, MT, NC NY, TX, OH & WA
- Product codes and lots
- 5.0.0.37, 5.0.1.11 and 5.0.2.35
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0079-2017 | Class II | Terminated | Voluntary: Firm initiated |