Z-0079-2017Class II

Radiation Therapy Treatment Planning System, Model 5.0

RaySearch Laboratories AB (publ) / initiated 2016-05-30 / Terminated / root cause: software

RAYSEARCH LABORATORIES AB began a recall of Radiation Therapy Treatment Planning System, Model 5.0 on . FDA classified it as Class II and gives the reason as "For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0079-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RaySearch Laboratories AB (publ)
Product
Radiation Therapy Treatment Planning System, Model 5.0
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics.
Root cause tag
software
FDA root cause
Software design
Action
On May 19th, 2016, RaySearch Laboratories distributed Field Safety Notice, Medical Device Correction notices and reply forms to their customers via email. Actions to be taken by the customer include: 1) Inspect all report templates that are intended to be used for plans with multiple beam sets. 2) Make sure that POI Dose statistics (Plan) is included immediately before ROI Dose statistics [Plan dose] in the Report content area. 3) Please educate planning staff and all users about this workaround. 4) Inspect your product and identify all installed units with the above software version number. Finally, complete and return the reply form via email to: [email protected] or by fax to: 888-501-7195 to confirm that you have read and understand the information. Customers with questions can contact David Hedfors via email: [email protected]
Quantity in commerce
491 units (187 domestically & 304 internationally)
Distribution
AZ, CA, FL, MT, NC NY, TX, OH & WA
Product codes and lots
5.0.0.37, 5.0.1.11 and 5.0.2.35

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0079-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0079-2017Class IITerminatedVoluntary: Firm initiated