Z-2689-2016Class II

A7 Anesthesia Delivery System, a device used to administer to a patient, continuously or intermittently, a...

Datascope Corp. / initiated 2016-06-15 / Terminated / root cause: manufacturing process

Mindray DS USA, Inc. dba Mindray North America began a recall of A7 Anesthesia Delivery System, a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a... on . FDA classified it as Class II and gives the reason as "Potential for a leak to occur on the back-up O2 and air e-size cylinder yokes on the A7 Anesthesia Delivery System". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2689-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
A7 Anesthesia Delivery System, a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. Part numbers 0632F-PA0000X (US domestic) and 0632B-00014 (international)
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
Potential for a leak to occur on the back-up O2 and air e-size cylinder yokes on the A7 Anesthesia Delivery System.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Mindray sent via certified mail with return receipt a recall letter dated June 15, 2016 to their affected customers. Mindray will replace the gasket on all e-size cylinder yokes on the affected A7 systems. Customer was instructed to contact their local Mindray Service Representative to arrange for this replacement. Customers can continue to use their A7 system while awaiting the replacement of the gasket(s). Customers can contact Ms. Diane Arpino, Director, Quality Operations and Regulatory Affairs via email to: [email protected] or via telephone to: (201)995-8407
Quantity in commerce
167 units (165 units - US) and (2 units - International)
Distribution
US Nationwide Distribution to AL, MA, MD, MN, MO, NE, NJ, OK, OR, PA and VA; and Canada
Product codes and lots
P/N 0632F-PA0000X - (Domestic) and 0632B-PA00014 - (International)   Serial numbers FR-5B000341, FR-5B000342, FR-5B000343, FR-5B000348, FR-5B000349, FR-51000148, FR-51000149, FR-51000150, FR-51000151, FR-51000152, FR-51000153, FR-51000154, FR-51000159 , FR-51000161, FR-51000162, FR-51000163, FR-51000164, FR-51000165, FR-51000166, FR-51000167, FR-51000168, FR-51000170, FR-51000171, FR-52000172, FR-52000173, FR-52000175, FR-52000176, FR-52000177, FR-52000178, FR-52000180, FR-52000181, FR-52000182, FR-52000183, FR-52000184, FR-52000185, FR-52000186, FR-53000187, FR-53000188, FR-53000189, FR-53000190, FR-53000191, FR-53000192, FR-53000193, FR-53000194, FR-53000197, FR-53000198, FR-53000199, FR-53000200, FR-53000203, FR-53000205, FR-53000206, FR-53000207, FR-53000209, FR-53000211, FR-53000212, FR-53000213, FR-53000214, FR-53000215, FR-53000216, FR-53000217, FR-54000218, FR-54000220, FR-54000221, FR-54000223, FR-54000224, FR-54000228, FR-54000229, FR-54000231, FR-54000233, FR-54000234, FR-54000235, FR-54000236, FR-54000237, FR-54000238, FR-54000239, FR-54000240, FR-54000241, FR-54000242, FR-54000243, FR-54000244, FR-56000245, FR-56000246, FR-56000247, FR-56000248, FR-56000249, FR-56000250, FR-56000251, FR-56000252, FR-56000253, FR-56000254, FR-56000255, FR-56000256, FR-56000257, FR-56000258, FR-56000259, FR-56000260, FR-56000261, FR-56000262, FR-56000265, FR-56000266, FR-56000267, FR-56000268, FR-57000269, FR-57000270, FR-57000271, FR-57000272, FR-57000274, FR-57000275, FR-57000276, FR-57000277, FR-57000278, FR-57000279, FR-57000280, FR-57000281, FR-57000282, FR-57000283, FR-57000284, FR-57000285, FR-57000286, FR-57000287, FR-57000288, FR-58000289, FR-58000290, FR-58000291, FR-58000292, FR-58000293, FR-58000294, FR-58000295, FR-58000296, FR-58000297, FR-58000298, FR-58000299, FR-58000300, FR-58000301, FR-58000302, FR-58000303, FR-58000304, FR-58000305, FR-58000306, FR-58000307, FR-58000308, FR-58000312, FR-58000313, FR-58000314, FR-58000315, FR-58000316, FR-58000317, FR-58000318, FR-58000319, FR-58000320, FR-58000321, FR-58000322, FR-58000323, FR-58000325, FR-59000330, FR-59000331, FR-59000333, FR-59000334, FR-59000335, FR-59000336, FR-59000338, and FR-59000339.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2689-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2689-2016Class IITerminatedVoluntary: Firm initiated