Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2630-2016Class II

120 V Neptune 3 Rover, Model Number: 0703-001-000

Stryker Corp. / initiated 2016-07-01 / Terminated

Stryker Instruments Div. of Stryker Corporation began a recall of 120 V Neptune 3 Rover, Model Number: 0703-001-000 on . FDA classified it as Class II and gives the reason as "Stryker initiated a voluntary recall of the 120 V Neptune 3 Rover on 07/01/2016, due to a potential for the device to experience system errors causing the device to shutdown if high levels of electromagnetic...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2630-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
120 V Neptune 3 Rover, Model Number: 0703-001-000
Product code
JCX Apparatus, Suction, Ward Use, Portable, Ac-Powered
Reason for recall
Stryker initiated a voluntary recall of the 120 V Neptune 3 Rover on 07/01/2016, due to a potential for the device to experience system errors causing the device to shutdown if high levels of electromagnetic interference are present in the operating room.
FDA root cause
Under Investigation by firm
Action
Stryker Instruments initiated a voluntary recall of the Stryker Neptune 3 Rover via certified mail on 07/01/2016, due to the potential for the Neptune 3 Rover to experience system errors if there are high levels of electromagnetic interference in the operating room. 1. Immediately review the Recall Notification. 2. Confirm that the Rover serial numbers identified on the Business Reply Form (BRF) are at your facility. 3. A Stryker representative will facilitate the repair of your affected Neptune 3 Rover(s). Please coordinate the availability of the product to assure the repair can be implemented. 4. Please complete and return the Business Reply Form (BRF). Note: Your signature on the BRF indicates that you received and understand this Notification and have followed the instructions in the Notification. 5. Fax the completed, signed Business Reply Form to Stryker Instruments Regulatory Department, 866-521-2762, or scan and email a copy to [email protected]. 6. Stryker will document the repair of each unit by serial number. For questions regarding this recall, please contact Stryker Instruments: Kelly Jo Davis 269-389-2921 [email protected]
Quantity in commerce
78
Distribution
Domestic:MI, CA, ID, IA VA/DOD: None Foreign: None
Product codes and lots
120 V Neptune 3 Rover, Model Number: 0703-001-000; Serial Numbers: 1606300523, 1610606263, 1610903973, 1610904293, 1612303683, 1612303823, 1612303963, 1606300553, 1610606273, 1610904063, 1610904313, 1612303703, 1612303833, 1612303973, 1606300583, 1610606303, 1610904083 , 1610904343, 1612303713, 1612303843 , 1612303993, 1606300593, 1610606323, 1610904103, 1610904353, 1612303723, 1612303853 , 1613106443, 1606300613, 1610606353, 1610904123, 1610904363 , 1612303733, 1612303863, 1613106463, 1606300623, 1610619473, 1610904163, 1610904373 , 1612303743, 1612303873, 1613106583, 1606300643, 1610619483, 1610904173, 1610904403, 1612303763 , 1612303883 , 1613201233, 1606300653, 1610619543, 1610904183, 1612303633, 1612303783, 1612303903, 1613201243 , 1607503463 , 1610619573 , 1610904193, 1612303643, 1612303793, 1612303913, 1613201303,  1607503493, 1610619603 , 1610904203, 1612303663, 1612303813, 612303933, 1613201323, 1610409473, 1610903893, 1610904213, 1612303673

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2630-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2630-2016Class IITerminatedVoluntary: Firm initiated