Merge PACS software. The firm name on the label is Merge Healthcare, Inc., Hartland, WI. Intended to create...
Merge Healthcare, Inc. began a recall of Merge PACS software. The firm name on the label is Merge Healthcare, Inc., Hartland, WI. Intended to create and display two-dimensional and three-dimensional... on . FDA classified it as Class II and gives the reason as "When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2363-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- Merge PACS software. The firm name on the label is Merge Healthcare, Inc., Hartland, WI. Intended to create and display two-dimensional and three-dimensional images of anatomy from a series of digitally acquired images.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm's customer support notified customers via phone call beginning 2/5/2016 to inform them of the issue. Customers were informed that a software fix that corrects the issue was available. Customers were asked not to use the measurements on mammo images until upgraded. If customers want an upgrade, then someone would contact them to schedule the upgrade.
- Quantity in commerce
- 49 sites potentially have the affected versions
- Distribution
- Worldwide Distribution -- USA and Australia.
- Product codes and lots
- Software versions 6.0.2.0 MR2, 6.0.3.0 MR3, 6.0.3.1 MR3 CU1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2363-2016 | Class II | Terminated | Voluntary: Firm initiated |