Z-2363-2016Class II

Merge PACS software. The firm name on the label is Merge Healthcare, Inc., Hartland, WI. Intended to create...

Merge Healthcare Incorporated / initiated 2016-02-05 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of Merge PACS software. The firm name on the label is Merge Healthcare, Inc., Hartland, WI. Intended to create and display two-dimensional and three-dimensional... on . FDA classified it as Class II and gives the reason as "When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2363-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge PACS software. The firm name on the label is Merge Healthcare, Inc., Hartland, WI. Intended to create and display two-dimensional and three-dimensional images of anatomy from a series of digitally acquired images.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm's customer support notified customers via phone call beginning 2/5/2016 to inform them of the issue. Customers were informed that a software fix that corrects the issue was available. Customers were asked not to use the measurements on mammo images until upgraded. If customers want an upgrade, then someone would contact them to schedule the upgrade.
Quantity in commerce
49 sites potentially have the affected versions
Distribution
Worldwide Distribution -- USA and Australia.
Product codes and lots
Software versions 6.0.2.0 MR2, 6.0.3.0 MR3, 6.0.3.1 MR3 CU1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2363-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2363-2016Class IITerminatedVoluntary: Firm initiated