Z-2629-2016Class II
OEC 9800, fluoroscopic x-ray system
GE Healthcare / initiated 2016-07-22 / Terminated / root cause: software
GE OEC Medical Systems, Inc began a recall of OEC 9800, fluoroscopic x-ray system on . FDA classified it as Class II and gives the reason as "A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2629-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- OEC 9800, fluoroscopic x-ray system
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- On July 22, 2016 GE Healthcare Surgery began distributing an Urgent Medical Device Correction letter to customers via mail to inform them of the issue and provide instructions for immediate mitigation of the issue. The letter was sent to 3 titles at each facility affected: 1.) Facility Administrator, 2.) Director/Manager of Radiology and Radiologists, 3.) Radiology Department. Consignees instructured that local GE Healthcare Field Engineer will contact them to coordinate replacement of the affected monitor(s), and will ensure that system meets specifications.
- Quantity in commerce
- 47 total
- Distribution
- Distribution US nationwide and Korea.
- Product codes and lots
- monitor part numbers: 5448359-01, 5459219-01, or 5475662-03:89-3853, 82-1593, 8S-1935, 89-3356, 89-1323, 8S-2885, 82-0700, 8S-1708, 82-0420, 82-1998, 8S-2027, 82-7101-MH, 8S-2670, P6-0375-L, 8S-1614, 89-0846, 82-0576, 82-3256, 8S-2202, 89-0711,89-1927, 89-1453, 82-1504, 82-0794, 82-3494, 8S-3125, 8S-1845-RC, 89-3725, 8S-1592, 89-3129,82-1182, 82-0986, 8S-0199, 8S-3613, 82-3748, 82-2335, 8S-2889, 82-0596, 89-3020, 82-1515, 8S-0645, 8S-3614, 82-3925, 89-0416, 89-1363, 82-3258, 8S-0347
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2628-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2629-2016 | Class II | Terminated | Voluntary: Firm initiated |