Z-2675-2016Class II

STAPLER,IS4000; Model number 470298; General and Plastic Surgery: The Intuitive Surgical da Vinci EndoWrist...

Intuitive Surgical, Inc. / initiated 2016-07-14 / Terminated / root cause: manufacturing process

Intuitive Surgical, Inc. began a recall of STAPLER,IS4000; Model number 470298; General and Plastic Surgery: The Intuitive Surgical da Vinci EndoWrist Stapler 45, Stapler 45 Reloads and other Stapler... on . FDA classified it as Class II and gives the reason as "Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30 instruments having an interruption in shaft rotation and / or instrument jaws not opening". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2675-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
STAPLER,IS4000; Model number 470298; General and Plastic Surgery: The Intuitive Surgical da Vinci EndoWrist Stapler 45, Stapler 45 Reloads and other Stapler Accessories are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device can be used with staple line or tissue butressing material (natural or synthetic).
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30 instruments having an interruption in shaft rotation and / or instrument jaws not opening.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Intuitive sent a Stop Use e-mail to all affected sites on July 14, 2016 to inform users of the issue prior to the commencement of a recall. E-mail described the problem and asked that all surgeons who use the stapler be informed to immediately stop use. Formal letters will follow. Medical Device Recall letters were sent on July 21, 2016. Customers were instructed to return quarantined affected product whether used or unused to Intuitive Surgical. Customers should contact customer service to initiate the standard Return Material Authorization (RMA) process by phone or email: North and South America: 800-876-1310, option 3 (6AM to 5PM PST) Email: [email protected] Europe, Middle East, Asia and Africa Phone: 800-821-2020 or +41 21 821 2020 Email: [email protected] Customers were also instructed to complete the attached Return Acknowledgement Form and submit via email to Regulatory Compliance at [email protected] or by fax to +1 (408) 523-0619. For questions regarding this recall call 408-523-2100. 10-31-16 Update: Monthly status update states that an additional 93 instruments have been identified and an additional communication will be sent to these sites in Mid November 2016,
Quantity in commerce
875 instruments (827 a la carte + 48 in kits) total, all models
Distribution
Worldwide Distribution - US (nationwide) Internationally to Australia, Belgium, Denmark, France, Germany. Italy. Japan, Spain, Sweden, Switzerland, Taiwan, and Turkey,
Product codes and lots
All model 470298

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2675-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2675-2016Class IITerminatedVoluntary: Firm initiated
Z-2676-2016Class IITerminatedVoluntary: Firm initiated
Z-2677-2016Class IITerminatedVoluntary: Firm initiated