Z-2749-2016Class II

Pacemakers: Accolade MRI DR Model L311; Accolade MRI DR EL Model L331; and Essentio MRI DR Model L111. These...

Boston Scientific Corporation / initiated 2016-06-15 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Pacemakers: Accolade MRI DR Model L311; Accolade MRI DR EL Model L331; and Essentio MRI DR Model L111. These pulse generators provide bradycardia pacing and... on . FDA classified it as Class II and gives the reason as "Five Accolate MRI and Essentio MRI DR pulse generators with an alternate MICS module configuration were shipped to US customers and implanted. This alternate configuration did not receive FDA approval yet". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2749-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Pacemakers: Accolade MRI DR Model L311; Accolade MRI DR EL Model L331; and Essentio MRI DR Model L111. These pulse generators provide bradycardia pacing and adaptive rate pacing to detect and treat bradyarrhythmias.
Product code
LWP Implantable Pulse Generator, Pacemaker (Non-Crt)
Reason for recall
Five Accolate MRI and Essentio MRI DR pulse generators with an alternate MICS module configuration were shipped to US customers and implanted. This alternate configuration did not receive FDA approval yet.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
As we considered the value of a direct physician communication we have not identified any information that would be of value to the physician because device performance and behavior remain unchanged.
Quantity in commerce
5
Distribution
Distributed to the states of TX, NH, MA, PA, and MN.
Product codes and lots
serial numbers 215173, 215186, 215507, 215855, 215957, 216051, 218956, 710312, and 710541.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2749-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2749-2016Class IITerminatedVoluntary: Firm initiated