Brilliance CT Big Bore Oncology, Model number 728243
Philips Medical Systems (Cleveland) Inc began a recall of Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by... on . FDA classified it as Class II and gives the reason as "After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2653-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- Customers were sent a letter dated July 28, 2016 informing them of the problem, under what circumstances it can occur and the actions that the customer can take to minimize the effect of the problem. The letter also includes the actions planned by Philips to correct the problem. Philips plans on releasing a field change order (FCO) to correct the affected systems. If you need any further information or support concerning this issue, contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1 -800-722-9377).
- Quantity in commerce
- 33
- Distribution
- US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
- Product codes and lots
- 728243 Brilliance BigBore Oncology CT 7116 7611 7354 7063 7309 7169 7205 7016 7019 7043 7069 7073 7091 7112 7132 7163 7141 7194 7226 7325 7294 7350 7316 7480 7114 7210 7310 7438 7530 7186 7119 7042
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2652-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2653-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2654-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2655-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2656-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2657-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2658-2016 | Class II | Terminated | Voluntary: Firm initiated |