Z-2656-2016Class II

Brilliance CT 16 Power, Model number 728240

Philips North America LLC / initiated 2016-07-20 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer... on . FDA classified it as Class II and gives the reason as "After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2656-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
Root cause tag
software
FDA root cause
Software Design Change
Action
Customers were sent a letter dated July 28, 2016 informing them of the problem, under what circumstances it can occur and the actions that the customer can take to minimize the effect of the problem. The letter also includes the actions planned by Philips to correct the problem. Philips plans on releasing a field change order (FCO) to correct the affected systems. If you need any further information or support concerning this issue, contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1 -800-722-9377).
Quantity in commerce
50
Distribution
US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
Product codes and lots
728240 -- Brilliance CT 16 Power  3100 5043 5047 5280 5345 5390 5413 5430 5495 5499 5512 6007 6009 6013 6032 6041 6045 6049 6051 6058 6064 6066 6070 6071 6077 6080 6081 6084 6089 6095 6100 6106 6109 6113 6115 6116 6123 6124 6131 6132 6134 6141 6144 6155 6161 6162 5476 6098 6120 6067

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2656-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2652-2016Class IITerminatedVoluntary: Firm initiated
Z-2653-2016Class IITerminatedVoluntary: Firm initiated
Z-2654-2016Class IITerminatedVoluntary: Firm initiated
Z-2655-2016Class IITerminatedVoluntary: Firm initiated
Z-2656-2016Class IITerminatedVoluntary: Firm initiated
Z-2657-2016Class IITerminatedVoluntary: Firm initiated
Z-2658-2016Class IITerminatedVoluntary: Firm initiated