Ingenuity Core128, Model number 728323
Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity Core128, Model number 728323 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer... on . FDA classified it as Class II and gives the reason as "After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2658-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Ingenuity Core128, Model number 728323 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- Customers were sent a letter dated July 28, 2016 informing them of the problem, under what circumstances it can occur and the actions that the customer can take to minimize the effect of the problem. The letter also includes the actions planned by Philips to correct the problem. Philips plans on releasing a field change order (FCO) to correct the affected systems. If you need any further information or support concerning this issue, contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1 -800-722-9377).
- Quantity in commerce
- 3
- Distribution
- US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
- Product codes and lots
- 728323 -- Ingenuity Core128 320004 320053 320056
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2652-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2653-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2654-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2655-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2656-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2657-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2658-2016 | Class II | Terminated | Voluntary: Firm initiated |