Alcon AcrySof IQ IOL with ULTRASERT Delivery System
Alcon Research, Ltd. began a recall of Alcon AcrySof IQ IOL with ULTRASERT Delivery System Product Usage: The Alcon AcrySof IQ Intraocular Lens (IOL) is an acrylic foldable singlepiece posterior... on . FDA classified it as Class II and gives the reason as "The ULTRASERT Delivery System from certain lots have an inferior surface characteristic that could result in the IOL becoming lodged in the Delivery System". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2708-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alcon Laboratories
- Product
- Alcon AcrySof IQ IOL with ULTRASERT Delivery System Product Usage: The Alcon AcrySof IQ Intraocular Lens (IOL) is an acrylic foldable singlepiece posterior chamber lens for the replacement of the human crystalline lens in the visual correction of aphakia in adult patients following cataract surgery. This material is capable of being folded prior to insertion. The lens gently unfolds to a fullsize lens body following implantation. The lens has a biconvex optic with supporting haptics The AcrySof IQ IOLs are provided in the ULTRASERT Preloaded Delivery System for a convenient, controlled means to reliably place these lenses into the capsular bag.
- Product code
- HQL Intraocular Lens
- Reason for recall
- The ULTRASERT Delivery System from certain lots have an inferior surface characteristic that could result in the IOL becoming lodged in the Delivery System
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Alcon Research sent an Urgent Field Safety Notice letter dated July 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to check their inventory and segregate any product being recalled. Customers were asked to complete and return the Field Safety Notice Response Form. For questions or concerns contact Alcon.
- Quantity in commerce
- 371 units
- Distribution
- Armenia, Belgium, Bulgaria, Croatia, Finland, France, Germany, Hungary, Italy, Lithuania, Malaysia, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Slovenia, Spain, Sweden, Switzerland, Thailand, United Kingdom (Great Britain and Ireland), and United Arab Emirates. *** While this product is available in the US, these lots were not distributed to US consignees ***
- Product codes and lots
- Lot numbers: 12407022, 12407085, 12407087, 12409013, 12409024, 12409026, 12409028
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2708-2016 | Class II | Terminated | Voluntary: Firm initiated |