Z-2708-2016Class II

Alcon AcrySof IQ IOL with ULTRASERT Delivery System

Alcon Laboratories / initiated 2016-07-28 / Terminated / root cause: manufacturing process

Alcon Research, Ltd. began a recall of Alcon AcrySof IQ IOL with ULTRASERT Delivery System Product Usage: The Alcon AcrySof IQ Intraocular Lens (IOL) is an acrylic foldable singlepiece posterior... on . FDA classified it as Class II and gives the reason as "The ULTRASERT Delivery System from certain lots have an inferior surface characteristic that could result in the IOL becoming lodged in the Delivery System". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2708-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alcon Laboratories
Product
Alcon AcrySof IQ IOL with ULTRASERT Delivery System Product Usage: The Alcon AcrySof IQ Intraocular Lens (IOL) is an acrylic foldable singlepiece posterior chamber lens for the replacement of the human crystalline lens in the visual correction of aphakia in adult patients following cataract surgery. This material is capable of being folded prior to insertion. The lens gently unfolds to a fullsize lens body following implantation. The lens has a biconvex optic with supporting haptics The AcrySof IQ IOLs are provided in the ULTRASERT Preloaded Delivery System for a convenient, controlled means to reliably place these lenses into the capsular bag.
Product code
HQL Intraocular Lens
Reason for recall
The ULTRASERT Delivery System from certain lots have an inferior surface characteristic that could result in the IOL becoming lodged in the Delivery System
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Alcon Research sent an Urgent Field Safety Notice letter dated July 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to check their inventory and segregate any product being recalled. Customers were asked to complete and return the Field Safety Notice Response Form. For questions or concerns contact Alcon.
Quantity in commerce
371 units
Distribution
Armenia, Belgium, Bulgaria, Croatia, Finland, France, Germany, Hungary, Italy, Lithuania, Malaysia, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Slovenia, Spain, Sweden, Switzerland, Thailand, United Kingdom (Great Britain and Ireland), and United Arab Emirates. *** While this product is available in the US, these lots were not distributed to US consignees ***
Product codes and lots
Lot numbers: 12407022, 12407085, 12407087, 12409013, 12409024, 12409026, 12409028

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2708-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2708-2016Class IITerminatedVoluntary: Firm initiated