510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2834-2016Class II

LAUREATE World Phaco System

Alcon Laboratories / initiated 2016-07-28 / Terminated / root cause: labeling and UDI

Alcon Research, Ltd. began a recall of LAUREATE World Phaco System on . FDA classified it as Class II and gives the reason as "Steps 16-25 are missing from the LAUREATE Multipak FMS Directions for Use (DFU)". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2834-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alcon Laboratories
Product
LAUREATE World Phaco System
Product code
HQC Unit, Phacofragmentation
Reason for recall
Steps 16-25 are missing from the LAUREATE Multipak FMS Directions for Use (DFU)
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
All remaining inventory of the identified affected products within Alcon DCs will be supplied corrected directions for use material prior to any further shipments.
Quantity in commerce
8,646 UNITS
Distribution
Burundi, China, Iran, South Korea, Taiwan, Thailand, USA. *** The recalled product is not sold in the USA or its territories. However, donated units are in use by a reduced number of USA clients for Medical Missions outside the country.
Product codes and lots
Lot Numbers: 1658863H, 1805033H, 1805034H, 1817516H, 1817517H, 1826141H, 1826142H, 1857622H, 1857623H, 1868399H, 1868400H

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2834-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2834-2016Class IITerminatedVoluntary: Firm initiated