Z-2833-2016Class II
TriFoillmaging Triumph/Triumph II Research CT scanners
Northridge Tri-Modality Imaging / initiated 2016-05-24 / Terminated
Northridge Tri-Modality Imaging, began a recall of TriFoillmaging Triumph/Triumph II Research CT scanners on . FDA classified it as Class II and gives the reason as "The Firm has determined that several Triumph/Triumph II Research CT scanners produced by Northridge Trimodality Imaging, inc dba TriFoillmaging (or by Gamma Medica-Ideas, Inc and field upgraded with a replacement x-ray...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2833-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Northridge Tri-Modality Imaging
- Product
- TriFoillmaging Triumph/Triumph II Research CT scanners.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The Firm has determined that several Triumph/Triumph II Research CT scanners produced by Northridge Trimodality Imaging, inc dba TriFoillmaging (or by Gamma Medica-Ideas, Inc and field upgraded with a replacement x-ray tube) may be out of compliance with 21 CFR 1020.40 (Cabinet X-ray standards)
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- TriFoil lmaging planned action to bring defect into Compliance: 1. All repairs were performed at customer facilities. Customers were contacted by phone to arrange a convenient repair time. 2. TriFoil service engineers installed a collimator on each of the affected cabinet x-ray systems to bring their emission rates into compliance with the emission limit in the performance standard. 3. Full radiation surveys were performed, at maximum exposure settings, to verify each repair was effective. 4. All costs associated with the repair and follow-up testing were paid directly by TriFoil Imaging. CDRH approves the CAP and understands that the CAP was completed on August 29, 2016. For further questions please call (818) 709-2468.
- Quantity in commerce
- 3
- Distribution
- USA Distribution
- Product codes and lots
- n/a
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2833-2016 | Class II | Terminated | FDA Mandated |