Z-2765-2016Class II

SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-20-OB (9G Introducer...

Hologic, Inc. / initiated 2016-08-02 / Terminated / root cause: manufacturing process

Hologic, Inc began a recall of SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-20-OB (9G Introducer Localization Set (OB Model) for ATEC0914-20MR... on . FDA classified it as Class II and gives the reason as "Tip Breakage at the glue joint between the obturator body and tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2765-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hologic, Inc.
Product
SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-20-OB (9G Introducer Localization Set (OB Model) for ATEC0914-20MR Handpiece)
Product code
GDF Guide, Needle, Surgical
Reason for recall
Tip Breakage at the glue joint between the obturator body and tip
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Hologic sent an Important Product Recall letter to customers in the United States through Certified letter. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately examine their inventory and cease use of all affected recalled product. Customers were asked to complete the enclosed Recall Response Card indicating whether you have any of the recalled product on hand, as well as the quantity of product that you have in stock.
Quantity in commerce
1440 units
Distribution
US Nationwide Distribution
Product codes and lots
Lot Codes: 512501, 512502, 512504, 512588, 601501, 601535, 601536, 601587, 602501, 602502, 602503, 602504, 603522, 604557, 604558, 604591, 605501, 605502, 605543, 605547.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2765-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2765-2016Class IITerminatedVoluntary: Firm initiated
Z-2766-2016Class IITerminatedVoluntary: Firm initiated