Z-2766-2016Class II
SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-12-OB (9G Introducer...
Hologic, Inc. / initiated 2016-08-02 / Terminated / root cause: manufacturing process
Hologic, Inc began a recall of SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-12-OB (9G Introducer Localization Set (OB Model) for ATEC0914-12MR... on . FDA classified it as Class II and gives the reason as "Tip Breakage at the glue joint between the obturator body and tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2766-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hologic, Inc.
- Product
- SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-12-OB (9G Introducer Localization Set (OB Model) for ATEC0914-12MR Handpiece Product Usage: The Sure Sight Obturator Introducer Localization System (ILS-OB) is a Hologic accessory kit. It is sold as an independent product for use with the ATEC MRI hand pieces. This sterile accessory kit is available in two lengths, is sold only in the US and has a 6 month shelf life. It contains a single use introducer stylet, introducer sheath, needle guide and localizing obturator
- Product code
- GDF Guide, Needle, Surgical
- Reason for recall
- Tip Breakage at the glue joint between the obturator body and tip
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Hologic sent an Important Product Recall letter to customers in the United States through Certified letter. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately examine their inventory and cease use of all affected recalled product. Customers were asked to complete the enclosed Recall Response Card indicating whether you have any of the recalled product on hand, as well as the quantity of product that you have in stock.
- Quantity in commerce
- 535 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Lot Codes: 512501, 512502, 512504, 512588, 601501, 601535, 601536, 601587, 602501, 602502, 602503, 602504, 603522, 604557, 604558, 604591, 605501, 605502, 605543, 605547.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2765-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2766-2016 | Class II | Terminated | Voluntary: Firm initiated |