Z-2625-2016Class II

SOMATOM Definition Flash with software version VA48A-SP2; Model # 10590000, computed tomography x-ray system

Siemens Medical Solutions USA, Inc. / initiated 2016-07-13 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Definition Flash with software version VA48A-SP2; Model # 10590000, computed tomography x-ray system on . FDA classified it as Class II and gives the reason as "Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2625-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Definition Flash with software version VA48A-SP2; Model # 10590000, computed tomography x-ray system
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Siemens mailed to customers an Important Customer Safety Notice to inform about a possible malfunction when using the affected systems and what actions to take to eliminate the problem. Also, Siemens informed their customers that they were currently developing a new firmware to eliminate the internal communication error, which should be released in August 2016. Customers were asked to contact their service organization at 800-888-7436, if they had any questions.
Quantity in commerce
2 systems
Distribution
Distributed to: MI, NY, CA, KY, ND, NE
Product codes and lots
Model # 10590000 Serial # 83398, and 83387

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2625-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2624-2016Class IITerminatedVoluntary: Firm initiated
Z-2625-2016Class IITerminatedVoluntary: Firm initiated
Z-2626-2016Class IITerminatedVoluntary: Firm initiated