SOMATOM Definition Edge with software version VA48A-SP2; Model # 8098027 computed tomography x-ray system
Siemens Medical Solutions USA, Inc began a recall of SOMATOM Definition Edge with software version VA48A-SP2; Model # 8098027 computed tomography x-ray system on . FDA classified it as Class II and gives the reason as "Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2626-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- SOMATOM Definition Edge with software version VA48A-SP2; Model # 8098027 computed tomography x-ray system
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.
- Root cause tag
- software
- FDA root cause
- Software in the Use Environment
- Action
- Siemens mailed to customers an Important Customer Safety Notice to inform about a possible malfunction when using the affected systems and what actions to take to eliminate the problem. Also, Siemens informed their customers that they were currently developing a new firmware to eliminate the internal communication error, which should be released in August 2016. Customers were asked to contact their service organization at 800-888-7436, if they had any questions.
- Quantity in commerce
- 4 systems
- Distribution
- Distributed to: MI, NY, CA, KY, ND, NE
- Product codes and lots
- Model # 8098027 Serial # 96020, 96047, 96011, and 96029
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2624-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2625-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2626-2016 | Class II | Terminated | Voluntary: Firm initiated |