510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2551-2016Class II

Merge Unity Z3D software. Merge Unity PACS (formerly DR Systems PACS) is a medical image and information...

Merge Healthcare Incorporated / initiated 2015-07-23 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of Merge Unity Z3D software. Merge Unity PACS (formerly DR Systems PACS) is a medical image and information management system that allows viewing, selection... on . FDA classified it as Class II and gives the reason as "The software is unable to accurately determine the calcium score of scans with a slice thickness not equal to 3 mm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2551-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge Unity Z3D software. Merge Unity PACS (formerly DR Systems PACS) is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
The software is unable to accurately determine the calcium score of scans with a slice thickness not equal to 3 mm.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Merge, sent a "canned" email on 7/23/2015 to its customers which also contained a personal message. The recalling firm issued a second letter entitled "URGENT: MEDICAL DEVICE RECALL" dated 1/27/2016 via e-mail on 1/30/2016. The letters described the product, problem and actions to be taken. The letter informed customers that an upgrade is available. The customers were instructed to call Merge Customer Service at (877) 741-5369 or send an email to [email protected] to schedule the upgrade-Reference this recall #2016-14. If you decide to decline this maintenance release, complete the attached form and return to Merge Healthcare via [email protected]. Alternatively, complete the online form: http://www.merge.com/Landing-Pages/Customer-Upgrade-Decline-Form.aspx no later than February 22, 2016. Ensure that all users of the product are provided with this notification. If you have further distributed this, identify your customers and notify them at once of this product recall and provide a copy of the recall notification letter. If you have any additional questions, call Merge Customer Service at (877) 741-5369 or [email protected].
Quantity in commerce
9 sites potentially have the affected version
Distribution
US Distribution to states of: CA, PA, TX, and MT.
Product codes and lots
Version 11.0

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2551-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2551-2016Class IITerminatedVoluntary: Firm initiated