Merge RadSuite software. Radiological image processing system
Merge Healthcare, Inc. began a recall of Merge RadSuite software. Radiological image processing system on . FDA classified it as Class II and gives the reason as "When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or studies for another...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2715-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- Merge RadSuite software. Radiological image processing system.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or studies for another patient.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued letters dated 4/3/2015 via email on 4/9/2015 notifying the customer of the issues.
- Quantity in commerce
- 10 sites have the affected software that is configured with the specific conditions listed in Code Information
- Distribution
- Distribution was made to medical facilities in AL, MI, MO, PA, TN, and TX.
- Product codes and lots
- RadSuite Versions: 5.30.8, 5.35.3, 5.35.4, 5.35.5, 8.30.7.3, 8.30.7.4, 8.30.7.5, 8.30.7.7 are affected, BUT ONLY if they are configured with all of the following conditions: (1) Must have IPID enabled; (2) Must use IPID morphers (3) Must Define a Defaultvalue in the IPID Morpher; (4) Must use the Aggregating Morpher; (5) Must not set OverwritelfPresent=false; and (6) Must be on an internal store (this is data that is already present on EA being stored to Radsuite, this excludes direct DICOM stores to the EA).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2715-2016 | Class II | Terminated | Voluntary: Firm initiated |