Z-2715-2016Class II

Merge RadSuite software. Radiological image processing system

Merge Healthcare Incorporated / initiated 2015-04-09 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of Merge RadSuite software. Radiological image processing system on . FDA classified it as Class II and gives the reason as "When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or studies for another...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2715-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge RadSuite software. Radiological image processing system.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or studies for another patient.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued letters dated 4/3/2015 via email on 4/9/2015 notifying the customer of the issues.
Quantity in commerce
10 sites have the affected software that is configured with the specific conditions listed in Code Information
Distribution
Distribution was made to medical facilities in AL, MI, MO, PA, TN, and TX.
Product codes and lots
RadSuite Versions: 5.30.8, 5.35.3, 5.35.4, 5.35.5, 8.30.7.3, 8.30.7.4, 8.30.7.5, 8.30.7.7 are affected, BUT ONLY if they are configured with all of the following conditions: (1) Must have IPID enabled;  (2) Must use IPID morphers  (3) Must Define a Defaultvalue in the IPID Morpher;  (4) Must use the Aggregating Morpher;  (5) Must not set OverwritelfPresent=false; and  (6) Must be on an internal store (this is data that is already present on EA being stored to Radsuite, this excludes direct DICOM stores to the EA).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2715-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2715-2016Class IITerminatedVoluntary: Firm initiated