Z-2796-2016Class II
RadSuite software. The firm name on the label is Merge Healthcare, Hartland, WI
Merge Healthcare Incorporated / initiated 2013-05-06 / Terminated / root cause: software
Merge Healthcare, Inc. began a recall of RadSuite software. The firm name on the label is Merge Healthcare, Hartland, WI on . FDA classified it as Class II and gives the reason as "Multiple studies were missing at a customer location that never made it to the archive before watermarking". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2796-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- RadSuite software. The firm name on the label is Merge Healthcare, Hartland, WI.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Multiple studies were missing at a customer location that never made it to the archive before watermarking
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm first issued a letter via email dated May 6, 2013. A letter dated January 28, 2016, was also issued via email. Both letters informed the customer a software upgrade was available and of the issue and makes recommendations about using the software. They are also informed an upgrade is available to correct the issue. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. In addition, the customer is requested to call Customer Service or send them an email to schedule the upgrade. If the customer chooses not to upgrade, they are requested to complete the attached form or go online and complete it. For further questions please call (800) 724-5970 Ext. 8027.
- Quantity in commerce
- 45 sites potentially have the affected versions
- Distribution
- Distribution was nationwide to medical facilities. Government distribution was made to LA and OK.
- Product codes and lots
- This potential patient safety issue is only present in the following versions when a reverse workflow is used: 2004-R6.x, 5.10.x, 5.20, 5.25, 5.30.1, 5.30.2, 5.30.2 SB0344, 5.30.3, 5.30.4, 5.30.4 HF0418E, 5.30.5, 5.30.5 HF0404E, 5.30.5 HF0413E, 5.30.5 HF390E, 5.30.5 SB0368, 5.30.6, 5.30.6 HF0405E, 5.30.6 HF0406E, 5.30.7, 5.30.7 HF0429E, 5.30.7 HF0441E, 5.30.7 HF0446E, 5.30.7 HF0470E, 5.30.7 HF0473E, 5.30.8, 5.30.8 HF0477E, 5.30.8 HF0490E, 5.30.8 HF0495 SB0508E, 5.30.8 HF0495E, 5.30.8 HF0495E SB504E, 5.30.8 HF0550, 5.30.8 SB0485E, 5.30.8 SB0486E, 5.30.X SB0381E, 5.35.1, 5.35.1 HF0487E, 5.35.1 HF0489E, 5.35.2, 5.35.3, 5.35.4, 5.35.4 HF0513, 5.35.4 HF0518, 5.35.4 HF0528, 5.35.4 HF0531, 5.35.4 HF0535, 5.35.4 HF0546, 8.30.0. NOTE: The standard workflow is not impacted. .
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2796-2016 | Class II | Terminated | Voluntary: Firm initiated |