Z-2763-2016Class II

Neural Signal Amplifier. Part Numbers: 4208, 5703, 5749, 5748, 5747

Blackrock Microsystems / initiated 2016-08-10 / Terminated / root cause: labeling and UDI

Blackrock Microsystems, LLC began a recall of Neural Signal Amplifier. Part Numbers: 4208, 5703, 5749, 5748, 5747. Product Usage: The Blackrock NeuroPort Biopotential Signal Processing System supports... on . FDA classified it as Class II and gives the reason as "Blackrock Microsystems, LLC announces a voluntary field action for the Blackrock NeuroPort Biopotential Signal Processing System because the Neural Signal Amplifier and the Patient Cable are incorrectly labeled as Type...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2763-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Blackrock Microsystems
Product
Neural Signal Amplifier. Part Numbers: 4208, 5703, 5749, 5748, 5747. Product Usage: The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes. Biopotential signals include: Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG), electrooculography (EOG) and Evoked Potential (EP)
Product code
GWL Amplifier, Physiological Signal
Reason for recall
Blackrock Microsystems, LLC announces a voluntary field action for the Blackrock NeuroPort Biopotential Signal Processing System because the Neural Signal Amplifier and the Patient Cable are incorrectly labeled as Type CF Applied Parts.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
Blackrock Microsystems sent an Urgent Medical Device letter dated August 12, 2016 via email. The letter identified the affected product, problem and actions to be taken. Customers were instructed not use the amplifier in application requiring a type CF applied part; perform a physical count of the inventory of amplifiers and patient cables with the part numbers listed. For questions regarding this Correction call 801-582- 5533, ext. 240.
Quantity in commerce
81 total amplifiers
Distribution
Worldwide Distribution - US Nationwide and Canada,
Product codes and lots
Part Number 4208: Serial Numbers 0014 and 0016 Only Part Number 5703: All Serial Numbers  Part Number 5748: All Serial Numbers  Part Number 5747: All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2763-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2763-2016Class IITerminatedVoluntary: Firm initiated
Z-2764-2016Class IITerminatedVoluntary: Firm initiated