3I OSSEOTITE CERTAIN DENTAL IMPLANTS, 4x11.5 mm, Sterile, Rx only

ZimVie US Corp LLC / initiated 2016-08-08 / Terminated / root cause: packaging

Biomet 3i, LLC began a recall of 3I OSSEOTITE CERTAIN DENTAL IMPLANTS, 4x11.5 mm, Sterile, Rx only on . FDA classified it as Class II and gives the reason as "Product was incorrectly packaged". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2845-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
3I OSSEOTITE CERTAIN DENTAL IMPLANTS, 4x11.5 mm, Sterile, Rx only.
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
Product was incorrectly packaged.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
An Urgent Medical Device Removal Notice is being sent to Customers August 12, 2016, notifying them of the packaging of OSSEOTITE Certain Implant (4x11.5mm) in T3 Implant boxes. This recall is specific to Lot No. 2016031461. The material shipped within the box was described as being correct ,but only the box it was shipped in was incorrect. Customers were notified that there are no immediate or long term health consequences. In the notification letter, customers are instructed to complete the Certificate of Acknowledgement, and to return it via email ([email protected]) if no product is on hand or to return the certificate with product by shipping materials to: Biomet 3i - Post-Market Returns. Replacement product will be sent to customers upon return. Customers were also advised to keep a copy of their Certificate of Acknowledgement. Further questions should be directed to 1-800 443-8166 between the hours of 8:00 AM and 6:00 PM, Monday thru Friday. For further questions, please call 1-800 342-5454.
Quantity in commerce
92 units
Distribution
US Distributions to the states of : CA and FL., and Worldwide to : Colombia, France, Germany, Italy, and Spain.
Product codes and lots
Model No: IOSS411; Lot No: 2016031461; UDI: (01)00844868007098(17)210401(10)2016031461.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2845-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2845-2016Class IITerminatedVoluntary: Firm initiated