Z-2802-2016Class II

RAPIDLab 1260 Blood Gas Analyzer Siemens Material Number (SMN): 10321846, 10491394

Siemens Healthcare Diagnostics Inc. / initiated 2016-08-11 / Terminated / root cause: software

Siemens Healthcare Diagnostics Inc began a recall of RAPIDLab 1260 Blood Gas Analyzer Siemens Material Number (SMN): 10321846, 10491394 on . FDA classified it as Class II and gives the reason as "There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2802-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
RAPIDLab 1260 Blood Gas Analyzer Siemens Material Number (SMN): 10321846, 10491394
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system
Root cause tag
software
FDA root cause
Software design
Action
Siemens Healthcare issued Urgent Field Safety Notice to (US) customers via Federal Express. . All affected customers outside the US provided a copy of an Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. This notice informs customers of the issue, Actions to be Taken by the Customer and to return the response form. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative
Quantity in commerce
114 units
Distribution
Nationwide Foreign: Albania Algeria Angola Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bosnia Herzeg. Botswana Brazil Brunei Bulgaria Canada Chile Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Fren.Polynesia Georgia Germany Greece Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Ivory Coast Japan Kazakhstan Kenya Kuwait Latvia Lebanon Lesotho Lithuania Luxembourg Macedonia Malaysia Mexico Netherlands New Caledonia New Zealand Norway P.R. China Peru Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Romania Russian Fed. Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Tunisia Turkey Turkmenistan U.A.E. United Kingdom Uruguay Vatikancity Vietnam
Product codes and lots
All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2802-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2797-2016Class IITerminatedVoluntary: Firm initiated
Z-2798-2016Class IITerminatedVoluntary: Firm initiated
Z-2799-2016Class IITerminatedVoluntary: Firm initiated
Z-2800-2016Class IITerminatedVoluntary: Firm initiated
Z-2801-2016Class IITerminatedVoluntary: Firm initiated
Z-2802-2016Class IITerminatedVoluntary: Firm initiated
Z-2803-2016Class IITerminatedVoluntary: Firm initiated