Z-2746-2016Class II

The Connecting Bracket (Fixture) Device (1002.65A0) is designed as a connecting element to support ancillary...

Maquet Cardiovascular Us Sales, LLC / initiated 2016-08-12 / Terminated / root cause: manufacturing process

Maquet Cardiovascular Us Sales, Llc began a recall of The Connecting Bracket (Fixture) Device (1002.65A0) is designed as a connecting element to support ancillary head rests on a Maquet operating room table... on . FDA classified it as Class II and gives the reason as "Post Market Surveillance cases reported to Maquet showing instances in which the welded joint between the square mount and the first fixable swivel joint of Maquet's Connecting Bracket (Fixture) Device is broken". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2746-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
The Connecting Bracket (Fixture) Device (1002.65A0) is designed as a connecting element to support ancillary head rests on a Maquet operating room table immediately before, during and after surgical interventions as well as for examination and treatment. The fixture is used to adapt a range of different head and horseshoe head rests with a central square mount
Product code
FQO Table, Operating-Room, Ac-Powered
Reason for recall
Post Market Surveillance cases reported to Maquet showing instances in which the welded joint between the square mount and the first fixable swivel joint of Maquet's Connecting Bracket (Fixture) Device is broken.
Root cause tag
manufacturing process
FDA root cause
Use error
Action
Maquet sent an "Urgent Product Recall Medical Device Correction Letter and Medical Device Correction Response Form" dated August 12, 2016 via UPS to their affected customers. Customers were advised to inspect all devices that include the fixture for damage before every use. Customers were asked to fill out and sign the enclosed Medical Device Correction Response Form and return it via email to [email protected] or fax to 1-973-909-9927. Customers with questions were advised to contact their MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383. For questions regarding this recall call 973-709-7487.
Quantity in commerce
19 units (US) and 1975 units (OUS)
Distribution
Nationwide Distribution to CA, FL, IA, IN, MO, UT and WI
Product codes and lots
Part Number/Catalog Number 1002.64A0

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2746-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2746-2016Class IITerminatedVoluntary: Firm initiated