DeRoyal (R) CYSTO TRAY PGYBK REF 89-3076.08, Rx Only, STERILE,EO; Custom surgical kit. O
DeRoyal Industries Inc began a recall of DeRoyal (R) CYSTO TRAY PGYBK REF 89-3076.08, Rx Only, STERILE,EO; Custom surgical kit. O on . FDA classified it as Class II and gives the reason as "The firm is recalling two custom surgical kits containing the BARD(TM) Toomey Irrigation Syringe 70cc with Resectoscope Tip, Catheter Tip, Luer Adapter and Cap. The BARD(TM) component is being recalled for a possible...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2679-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DeRoyal Industries, Inc.
- Product
- DeRoyal (R) CYSTO TRAY PGYBK REF 89-3076.08, Rx Only, STERILE,EO; Custom surgical kit. O,
- Product code
- LRO General Surgery Tray
- Reason for recall
- The firm is recalling two custom surgical kits containing the BARD(TM) Toomey Irrigation Syringe 70cc with Resectoscope Tip, Catheter Tip, Luer Adapter and Cap. The BARD(TM) component is being recalled for a possible slit defect on the package that may affect the component's sterility.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- DeRoyal issued their recall on 08/11/2016 to the end-user level by sending overnight letters to its direct accounts. The customers were asked to identify the affected trays, place them in quarantine and return them to DeRoyal for replacement or credit.
- Quantity in commerce
- 24 units
- Distribution
- Distributed in LA and TN.
- Product codes and lots
- Lot Number 40004006
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2678-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2679-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY