Z-2757-2016Class II

Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula...

LivaNova USA, Inc. / initiated 2016-08-12 / Terminated / root cause: manufacturing process

Sorin Group USA, Inc. began a recall of Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single... on . FDA classified it as Class II and gives the reason as "The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2757-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided for with a vinyl female luer connector bonded proximally and a vinyl cuff bonded distally. The cuff is pre-molded to its inflated configuration and is purposely positioned directly over a hole generated in the extruded tube for purposes of receiving the infused cardioplegia solution. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
LivaNova issued an Urgent Safety Alert letter via certified mail on August 12, 2016 to all affected customers. The letter instructed customers to check their inventory to determine if the Cannulae belongs to the product code and lots listed. If the customer does not have any of the products described in the communication, please complete the response form below and return it per the instructions indicated on the form; otherwise, please check the tip angle to determine if it is conforming to specifications. If specification is not met, please file a complaint through your normal process whether you decide to keep it and use it or return the product to Sorin Group; if you are not willing to use it, return the affected parts to Sorin Group USA. Please contact Customer Service 1-S00-650-2623 or email [email protected] to have a Return Material Authorization (RMA) issued. Complete the Customer Response Form attached to this letter to acknowledge receipt of this letter. Please return this form no later than September 15, 2016. For questions regarding this recall call 303-467-6306.
Quantity in commerce
1205 units worldwide
Distribution
Worldwide Distribution - US including IL, MO, TX. Internationally to Canada and Italy.
Product codes and lots
Lot Numbers: 1425900137, 1433900079, 1433900080, 1435300088.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2757-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2756-2016Class IITerminatedVoluntary: Firm initiated
Z-2757-2016Class IITerminatedVoluntary: Firm initiated