VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit), Unique Device...
Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit), Unique Device Identifier No. 10758750000340, Rx ONLY, IVD --- For... on . FDA classified it as Class II and gives the reason as "Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0092-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit), Unique Device Identifier No. 10758750000340, Rx ONLY, IVD --- For in vitro diagnostic use only.
- Product code
- KNK Acid, Uric, Uricase (Colorimetric)
- Reason for recall
- Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.
- FDA root cause
- Under Investigation by firm
- Action
- Ortho-Clinical Diagnostics sent an URGENT PRODUCT CORRECTION Letter (Ref. CL2016-152) dated August 11, 2016 via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to all customers who could order and use affected product to inform them that new interferant were identified for GLU, LAC, TRIG and URIC Slides which may cause biased results. Foreign affiliates were informed by e-mail on August 11, 2016, of the issue and instructed to notify their consignees of the issue. For questions, please contact our Ortho CareTM Technical Solutions Center at 1-800-421-3311. For questions regarding this recall call 908-218-8776.
- Quantity in commerce
- US: 18,749 sales units; Foreign: 94,928 sales units
- Distribution
- Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.
- Product codes and lots
- Current/In-date GENs 26, 27, 28, 29, 31
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0089-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0090-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0091-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0092-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
