HexaPOD evo RT System
Elekta, Inc. began a recall of HexaPOD evo RT System Product Usage: The intended use of the device is the control of accurate patient positioning with assistance of a 3D Tracking System in a... on . FDA classified it as Class II and gives the reason as "Potentially unrecognized incorrect position of the treatment couch in 3D workflow, i.e. the HexaPOD has not moved fully to the 3D position". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2691-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- HexaPOD evo RT System Product Usage: The intended use of the device is the control of accurate patient positioning with assistance of a 3D Tracking System in a radiotherapy environment.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Potentially unrecognized incorrect position of the treatment couch in 3D workflow, i.e. the HexaPOD has not moved fully to the 3D position.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Elekta sent an Important User Notice 618-03-303-025 to affected customers on 8/18/2016. The notice identified the affected product, problem, clinical impact, solution and actions to be taken. The Notice reminds users that the Interlock Check is not possible during 3D Workflow as iGUIDE is not in use. The notice reminds users that if iGUIDE is not being used - (A) Ensure the HexaPOD is at its pre-defined 3D position before treatment, and (B) Ensure the External Inhibit LED at the EnableSwitch Board is off. For questions contact your local Elekta office.
- Quantity in commerce
- 13
- Distribution
- Worldwide Distribution - US Nationwide in the states of LA, PA, WA, and the countries of: Australia, Denmark, France, India, Italy, and Japan.
- Product codes and lots
- iGuide 2.2.0
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2691-2016 | Class II | Terminated | Voluntary: Firm initiated |