BrightView XCT Upgrade Model 882454
Philips Medical Systems (Cleveland) Inc began a recall of BrightView XCT Upgrade Model 882454 on . FDA classified it as Class II and gives the reason as "The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2737-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- BrightView XCT Upgrade Model 882454
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm is informing the consignee of this issue through Customer Information Letter.
- Quantity in commerce
- 3 units
- Distribution
- Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
- Product codes and lots
- 11000021, 11000082, 11000111
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2735-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2736-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2737-2016 | Class II | Terminated | Voluntary: Firm initiated |