3.5mm Bowed Locking Compression Femur Plates, 14 hole Orthopediatrics PediLoc Locking Plate System comes in...
OrthoPediatrics Corp began a recall of 3.5mm Bowed Locking Compression Femur Plates, 14 hole Orthopediatrics PediLoc Locking Plate System comes in sets that offer the advantage of both conventional... on . FDA classified it as Class II and gives the reason as "OrthoPediatrics Corp.is voluntarily recalling multiple lots of the 3.5mm Bowed Locking Compression Femur 14-hole Plate and 18-hole Plate due to the devices being mislabeled". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2843-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthopediatrics Corp
- Product
- 3.5mm Bowed Locking Compression Femur Plates, 14 hole Orthopediatrics PediLoc Locking Plate System comes in sets that offer the advantage of both conventional and locking plate fixation devices and instrumentation in one system. Utilizing both locking and non- locking screws, PediLoc offers a construct that resists angular collapse while simultaneously acting as an effective aid to fracture reduction.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- OrthoPediatrics Corp.is voluntarily recalling multiple lots of the 3.5mm Bowed Locking Compression Femur 14-hole Plate and 18-hole Plate due to the devices being mislabeled.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Orthopediatrics initiated the recall of 3.5mm Bowed Locking Compression Femur Plates by distributing e-mails on 0July 27, 2016, followed by telephone calls to all customers on July 28, 2016. All distributors and direct representatives were sent an e-mail to provide them with a written notification of the voluntary recall and instruct them to quarantine all identified devices from the affected lot numbers immediately. Customers were instructed to quarantine recalled product and response via a hard copy Reply Form that was included in a hard copy of the Recall Notice. Each sales representative was also asked to contact Orthopediatrics Logistics Department to receive a Return Authorization Number for returning recalled product. in order to track their affected plates. If customers have any questions about this recall, they were instructed to reach out to the firm via phone (office: 574-267-0865) or email ([email protected]).
- Quantity in commerce
- 41
- Distribution
- Nationwide Distribution to AZ, CA, CO, DE, FL, GA, KY, MI, MN, MO, KY, RI, NY, OH, OR, RI, TN, TX, VA, WA Foreign: None VA/DOD: None
- Product codes and lots
- Product Number: 00-1050-3214 Lot Number: NM02341 Product Number: 00-1050-3214 Lot Number: NM02342
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2843-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2844-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.