Z-2789-2016Class II

Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10)...

Howmedica Osteonics Corp / initiated 2016-08-01 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10) The Scorpio Patella Assembly Instrument is used... on . FDA classified it as Class II and gives the reason as "Upon failure of the press-fit, the clamping sub components and pins could disassociate from the Scorpio Patella Assembly Instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2789-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10) The Scorpio Patella Assembly Instrument is used during total knee replacement surgery for the implantation of Scorpio Patella implant components. The device functions to fully seat the implant onto the bone and secure it in position while the surgeon removes any excess cement and when cement polymerization occurs.
Product code
MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
Reason for recall
Upon failure of the press-fit, the clamping sub components and pins could disassociate from the Scorpio Patella Assembly Instrument.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Stryker sent an "Urgent Medical Device Recall Notification" Letter and Recall Business Reply Form dated August 1, 2016 to their affected customers via email and also via UPS on August 1, 2016. Customers are asked to quarantine the product subject to this recall. Branches/Agencies are asked to return all affected products available at their location to: Stryker C/O Stericycle, Attn: RA 2016-098 - Event 8296. Hospital/Branches: Customers are asked to complete and sign the Business Reply Form and fax a copy within 5 business days to: 877-272-6141 or email to Stericycle [email protected]. Any questions, customers can call (201) 831-6693.
Quantity in commerce
2,039 units
Distribution
US Nationwide and Internationally
Product codes and lots
Catalog No. 3182-1000 Lot #'s: RD8E100, RD8A054TT, RD8A054T, RD8E099M, RD8E099A, RD8E099, RD8A054A, RD8E099D, RD8A053A, RD8A054, RD8A053T, RD7K062L, RD8A052A, RD8A052E, RD8A052, RD8A053, RD7E128D, RD5T290H, RD7E128T, RD5S262D, RD5T290W, RD7K062E, RD5S262X, RD7E128Y, RD7A076MA, RD6W102MA, RD7K062A, RD7K062, RD7E128H, RD7A076A, RD7E128A, RS7E128, RD6W103D, RD6V108J, RD6V108K, RD6V108H, RD7A076H, RD6V112M, RD7A074D, RD7A076J, RD6V112H, RD6W102m, RD5H286A, RD2V022, RD2K049, RD6W103, RD5H286J, RD2T037A, RD7A076M, RD6S136A, RD5H286, RD2V031C, RD5N361T, RD6V112A, RD4M068J, RD2V021C, RD7A076, RD6V112, RD4E226W, RD2V021, RD7A074H, RD6W102, RD4S220W, RD2V023X2, RD6W103A, RD6S136M, RD4M069D, RD2V023X3, RD6V108, RD5W148M, RD4S220, RD2V023, RD7A074, RD5M237K, RD4S220A, RD2V031A, RD7A074A, RD5T290X, RD4S220A1, RD2V023X1, RD6W103M, RD5T290A, RD4M069, RD2V031, RD5K116A, RD6S137A, RD4L128X1, RD2V031X2, RD5M236T, RD5W148A, RD4L128V, RD2V031X1, RD5K116T, RD6S137, RD4E226J, RD2N030K, RD3V002A, RD6S136, RD4M068< RD2T037, RD5M237, RD5W148X, RD4L128, RD2T037X, RD5K116N, RD5W148, RD4E226M, RD2N030E, RD5K116, RD5T290M, RD4E226L1, RD2N030S, RD5M236, RD5T290E, RD4C230L, RD2N030X, RD5H286L, RD5T290, RD4E226L, RD2N030XX, RD3T155L, RD5T290D, RD4E226, RD2N030, RD2V021E, RD5N361A, RD4C162L, RD2K068W, RD2V022W, RD5S305J, RD4C162, RD2N030X1, RD2V021S, RD5S305M, RD4C230, RD2M019E, RD2V022K, RD5N361M, RD4C230X, RD2M039, RD2V022E, RD5S305, RD3V002P, RD2M019H, RD2V021H, RD5S262A, RD3V002X1, RD2K049H, RD2V023E, RD5S262M1, RD3V002X2, RD2M003H, RD2T037W, RD5S262M, RD3V002D1, RD2M038H, RD2K068X1, RD5S262J, RD3V002, RD2K068B, RD2K068, RD5S262, RD3V002D, RD2M003, RD2K049R, RD5N361H, RD3V206D, RD2M038, RD2K049E, RD5M259M, RD3T160L1, RD2M019, RD2K017E, RD5N316, RD3V206X1, RD2K068S, RD2H056, RD5H286D, RD3T155L1, RD2K056S, RD2K056, RD5M259T, RD3T160L3, RD2K065S, RD2K065, RD5M236H, RD3T160, RD2K049S, RD2K017, RD5M259, RD3T160L2, RD2K068E, RD5M237L, RD3T160L, RD2H056E, RD5K116J, RD3V206, RD2K065E, RD4S220D, RD3T155 and RD2K056E

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2789-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2789-2016Class IITerminatedVoluntary: Firm initiated