Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Image processing system
Merge Healthcare, Inc. began a recall of Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Image processing system on . FDA classified it as Class II and gives the reason as "A report can be confirmed with the incorrect patient demographics resulting in it being saved under the incorrect patient record in the Electronic Medical Record (EMR)". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2709-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Image processing system.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- A report can be confirmed with the incorrect patient demographics resulting in it being saved under the incorrect patient record in the Electronic Medical Record (EMR).
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- MERGE sent an Urgent Medical Device Recall letter dated February 9, 2016, to all affected customers via email or via certified mail if they did not have an e-mail address for the customer. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. In addition, the customer is requested to call Customer Service or send them an email to schedule the upgrade. If the customer chooses not to upgrade, they are requested to complete the attached form or go online and complete it. For further questions, please call (800) 724-5970 ext. 8027.
- Quantity in commerce
- 116 customers potentially have the affected versions
- Distribution
- Nationwide Distribution
- Product codes and lots
- Versions 10.1 and 10.1.1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2709-2016 | Class II | Terminated | Voluntary: Firm initiated |