Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console
Sorin Group USA, Inc. began a recall of Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during... on . FDA classified it as Class II and gives the reason as "Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2866-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Sorin Group Deutschland GmbH created Field Correction Order (FCO) 2016-003 on August 11, 2016 to replace the affected power supply with a new unit. The customers will be contacted by the Field Service Representatives to schedule the replacement at the customer facilities. No direct action is required by the customer.
- Quantity in commerce
- 5264 Units Worldwide
- Distribution
- Worldwide Distribution - US Nationwide in the states of AZ, CA, CO, FL, HI, KS, MA, NC, TX, UT and the countries of: Great Brittan, Peru, Russia, India, Hungary, Thailand, Germany, Azerbaijan, Austria, Italy, Slovakia, Bahrain, and Malaysia.
- Product codes and lots
- Item No.: 48-30-00 Serial No: 28E24640, 28E24651, 28E24642, 28E24674, 28E24657, 28E24652, 28E24671, 28E24675, 28E24676, 28E24678, 28E24670 Item No 48-40-00 Serial No: 28E24644, 28E24641, 28E24648, 28E24661, 28E24643, 28E24650, 28E24677, 28E24659, 28E24658, 28E24672, 28E24666, 28E24660, 28E24665, 28E24645, 28E24646, 28E24667, 28E24668, 28E24669, 28E24679 Item No 48-50-00 Serial No: 28E24647, 28E24663, 28E24673, 28E24664 Item No 58-00-00 Serial No.: 28E24653, 28E24654, 28E24655, 28E24656
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2866-2016 | Class II | Terminated | Voluntary: Firm initiated |