Z-2866-2016Class II

Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console

LivaNova USA, Inc. / initiated 2016-08-25 / Terminated / root cause: manufacturing process

Sorin Group USA, Inc. began a recall of Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during... on . FDA classified it as Class II and gives the reason as "Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2866-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Sorin Group Deutschland GmbH created Field Correction Order (FCO) 2016-003 on August 11, 2016 to replace the affected power supply with a new unit. The customers will be contacted by the Field Service Representatives to schedule the replacement at the customer facilities. No direct action is required by the customer.
Quantity in commerce
5264 Units Worldwide
Distribution
Worldwide Distribution - US Nationwide in the states of AZ, CA, CO, FL, HI, KS, MA, NC, TX, UT and the countries of: Great Brittan, Peru, Russia, India, Hungary, Thailand, Germany, Azerbaijan, Austria, Italy, Slovakia, Bahrain, and Malaysia.
Product codes and lots
Item No.: 48-30-00 Serial No: 28E24640, 28E24651, 28E24642, 28E24674, 28E24657, 28E24652, 28E24671, 28E24675, 28E24676, 28E24678, 28E24670  Item No 48-40-00 Serial No: 28E24644, 28E24641, 28E24648, 28E24661, 28E24643, 28E24650, 28E24677, 28E24659, 28E24658, 28E24672, 28E24666, 28E24660, 28E24665, 28E24645, 28E24646, 28E24667, 28E24668, 28E24669, 28E24679  Item No 48-50-00 Serial No: 28E24647, 28E24663, 28E24673, 28E24664  Item No 58-00-00 Serial No.: 28E24653, 28E24654, 28E24655, 28E24656

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2866-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2866-2016Class IITerminatedVoluntary: Firm initiated