510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2855-2016Class II

SENSAR Foldable IOL with OptiEdge Design, Model No. AR40e The product is indicated for implantation for the...

Advanced Medical Optics, Inc. / initiated 2016-08-26 / Terminated / root cause: manufacturing process

Abbott Medical Optics Inc. (AMO) began a recall of SENSAR Foldable IOL with OptiEdge Design, Model No. AR40e The product is indicated for implantation for the visual correction of aphakia in adult patients in... on . FDA classified it as Class II and gives the reason as "Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2855-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Medical Optics, Inc.
Product
SENSAR Foldable IOL with OptiEdge Design, Model No. AR40e The product is indicated for implantation for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction. The lens is intended to be placed in the capsular bag.
Product code
HQL Intraocular Lens
Reason for recall
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Abbott Medical Optics, Inc. sent an Urgent Field Safety Notice to all affected customers to inform them that Abbott Medical Optics (AMO) is recalling 737 IOLs due to detection of an inspection equipment malfunction that resulted in a total of 737 distributed lenses not being fully checked prior to release. The letter informs customers that this malfunction may result in the release of mislabeled IOLs. The letter informs the customers of the actions to be taken. Customers are instructed to complete Customer Reply Form and fax it to AMO Quality Assurance at 714-247-4510 or email to [email protected]. Customers with any questions related to the IOL return process are instructed to contact an AMO Customer Service Representative at 877-266-4543. Customers with product complaints or adverse events are instructed to inform AMO by calling 877-266-4543.
Quantity in commerce
84 units
Distribution
Worldwide Distribution - US nationwide, PR, and Internationally to Canada, Latin America, Europe, Africa, Middle East, and Asia Pacific.
Product codes and lots
Serial No. 5626941507 5626961507 5626981507 5627001507 2003061505 2004131506 2015871503 3848581509 5626621507 5626641507 5626651507 5626661507 5626671507 5626681507 5626691507 5626701507 5626711507 5626721507 5626731507 5626741507 5626751507 5626761507 5626771507 5626791507 5626801507 5626811507 5626821507 5626831507 5626841507 5626851507 5626861507 5626871507 5626881507 5626891507 5626901507 5626911507 5626921507 5626931507 5626951507 5626971507 5627011507 5627021507 5627031507 5627041507 5627051507 5627061507 5627071507 5627081507 5627091507 5627101507 5627111507 5627121507 5627131507 5627141507 5627151507 5627161507 5627171507 5627181507 5627191507 5627201507 5627211507 5627221507

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2855-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2855-2016Class IITerminatedVoluntary: Firm initiated
Z-2856-2016Class IITerminatedVoluntary: Firm initiated
Z-2857-2016Class IITerminatedVoluntary: Firm initiated
Z-2858-2016Class IITerminatedVoluntary: Firm initiated
Z-2859-2016Class IITerminatedVoluntary: Firm initiated
Z-2860-2016Class IITerminatedVoluntary: Firm initiated
Z-2861-2016Class IITerminatedVoluntary: Firm initiated
Z-2862-2016Class IITerminatedVoluntary: Firm initiated
Z-2863-2016Class IITerminatedVoluntary: Firm initiated
Z-2864-2016Class IITerminatedVoluntary: Firm initiated