Z-2857-2016Class II

TECNIS Symfony Extended Range of Vision IOL, Model No. ZXR00 The product is indicated primary implantation...

Advanced Medical Optics, Inc. / initiated 2016-08-26 / Terminated / root cause: manufacturing process

Abbott Medical Optics Inc. (AMO) began a recall of TECNIS Symfony Extended Range of Vision IOL, Model No. ZXR00 The product is indicated primary implantation for the visual correction of aphakia in adult... on . FDA classified it as Class II and gives the reason as "Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2857-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Medical Optics, Inc.
Product
TECNIS Symfony Extended Range of Vision IOL, Model No. ZXR00 The product is indicated primary implantation for the visual correction of aphakia in adult patients with and without presbyopia in whom a cataractous lens has been removed by extracapsular cataract extraction, and aphakia following refractive lensectomy in presbyopic adults, who desire useful vision over a continuous range of distances including far, intermediate and near, and increased spectacle independence. The lens is intended to be placed in the capsular bag.
Product code
HQL Intraocular Lens
Reason for recall
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Abbott Medical Optics, Inc. sent an Urgent Field Safety Notice to all affected customers to inform them that Abbott Medical Optics (AMO) is recalling 737 IOLs due to detection of an inspection equipment malfunction that resulted in a total of 737 distributed lenses not being fully checked prior to release. The letter informs customers that this malfunction may result in the release of mislabeled IOLs. The letter informs the customers of the actions to be taken. Customers are instructed to complete Customer Reply Form and fax it to AMO Quality Assurance at 714-247-4510 or email to [email protected]. Customers with any questions related to the IOL return process are instructed to contact an AMO Customer Service Representative at 877-266-4543. Customers with product complaints or adverse events are instructed to inform AMO by calling 877-266-4543.
Quantity in commerce
10 units
Distribution
Worldwide Distribution - US nationwide, PR, and Internationally to Canada, Latin America, Europe, Africa, Middle East, and Asia Pacific.
Product codes and lots
Serial No. 8618391605 8618441605 8231911605 8231921605 8231931605 8231961605 8496921508 8496931508 8496941508 8496951508

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2857-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2855-2016Class IITerminatedVoluntary: Firm initiated
Z-2856-2016Class IITerminatedVoluntary: Firm initiated
Z-2857-2016Class IITerminatedVoluntary: Firm initiated
Z-2858-2016Class IITerminatedVoluntary: Firm initiated
Z-2859-2016Class IITerminatedVoluntary: Firm initiated
Z-2860-2016Class IITerminatedVoluntary: Firm initiated
Z-2861-2016Class IITerminatedVoluntary: Firm initiated
Z-2862-2016Class IITerminatedVoluntary: Firm initiated
Z-2863-2016Class IITerminatedVoluntary: Firm initiated
Z-2864-2016Class IITerminatedVoluntary: Firm initiated