IMMULITE 2000/IMMULITE 2000XPi Systems CMM CMV IgM Assay, REF/Catalog Number L2KCM2(D) - 200 tests (US), SMN...
Siemens Healthcare Diagnostics, Inc. began a recall of IMMULITE 2000/IMMULITE 2000XPi Systems CMM CMV IgM Assay, REF/Catalog Number L2KCM2(D) - 200 tests (US), SMN 10370302 (US), UDI 00630414985725 (US); and... on . FDA classified it as Class II and gives the reason as "Multiple customers reported seeing an increase in imprecision with some patients resulting in high % CVs with the IMMULITE 2000/2000XPi CMV (IgM) assay starting with reagent lots 259 and above". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0320-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- IMMULITE 2000/IMMULITE 2000XPi Systems CMM CMV IgM Assay, REF/Catalog Number L2KCM2(D) - 200 tests (US), SMN 10370302 (US), UDI 00630414985725 (US); and REF/Catalog Number L2KCM2 - 200 tests (OUS), SMN 10371320 (OUS), UDI 00630414961163 (OUS); IVD --- Siemens Healthcare Diagnostics Products Ltd., Llanberis, Gwynedd, LL55 4EL UK For in vitro diagnostic use with IMMULITE 2000 analyzers for the qualitative detection of IgM antibodies to cytomegalovirus (CMV) in human serum or plasma (EDTA or heparinized), as an aid in the diagnosis of current and recent CMV infection in individuals with signs and symptoms of CMV infection or clinical suspicion of CMV infection. This assay is not FDA cleared or approved for use in testing (screening) blood or plasma donors, neonatal screening, or for use at point-of-care facilities.
- Product code
- LKQ Antibody Igm,If, Cytomegalovirus Virus
- Reason for recall
- Multiple customers reported seeing an increase in imprecision with some patients resulting in high % CVs with the IMMULITE 2000/2000XPi CMV (IgM) assay starting with reagent lots 259 and above.
- FDA root cause
- Under Investigation by firm
- Action
- The firm, Siemens, issued an "Urgent Medical Device Recall" Letter (IMC16-22.A.US, dated 8/30/2016) in the US and an "Urgent Field Safety Notice" (IMC16-22.A.OUS, dated August 2016) Outside the US (OUS) starting on 8/30/2016 The notices instructed customers to discontinue use of and discard the affected kit lots listed in the customer notice; review your inventory of the products to determine your laboratory's replacement needs; review the UMDR and or the UFSN with their Medical Director, In the US, the method of distribution is FedEx and Outside the US the distribution is determined at the country level; complete and return the Field Correction Effectiveness Check Form via Fax to: Customer Care Center at (312) 275-7795 within 30 days, and retain this letter with your laboratory records, and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative.
- Quantity in commerce
- US: 326 kits; Foreign: 9633 kits
- Distribution
- Worldwide distribution- US (Nationwide) and countries of: United Arab Emirates, Afghanistan, Angola, Argentina, Australia, Bangladesh, Belgium, Bulgaria, Bahrain, Bolivia, Brazil, Belarus, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Dominican Republic, Algeria, Ecuador, Egypt, Spain, France, Great Britain, Greece, Guatemala, Honduras, Hungary, India, Iraq, Italy, Jordan, Canary Islands, Kyrgyzstan, Rep. Korea (S), Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macedonia, Mexico, Malaysia, Nicaragua, Netherlands, Norway, Nepal, Peru, Pakistan, Poland, Portugal, Paraguay, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, El Salvador, Syria, Turkmenistan, Tunisia, Turkey, Uruguay, Uzbekistan, Vietnam, Serbia,and South Africa.
- Product codes and lots
- UDI 00630414985725/Catalog # L2KCM2(D): Kit lots D255, D256, D257, D258, D259, D260, D261, D262, D263, D264, D266; UDI 00630414961163/Catalog # L2KCM2: Kit lots 255, 256, 257, 258, 259, 260, 261, 262, 263, 264, 266, 267
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0319-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0320-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.