Z-2912-2016Class III

Siemens RAPIDPoint 400 Measurement Cartridge Siemens SMN numbers:10329756, 10341161

Siemens Healthcare Diagnostics Inc. / initiated 2016-09-02 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics Inc began a recall of Siemens RAPIDPoint 400 Measurement Cartridge Siemens SMN numbers:10329756, 10341161 on . FDA classified it as Class III and gives the reason as "Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2912-2016
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens RAPIDPoint 400 Measurement Cartridge Siemens SMN numbers:10329756, 10341161
Product code
JFS Coulometric, Chloride
Reason for recall
Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Siemens Healthcare issued An Urgent Field Safety Notice via e-mail to affected Siemens Healthcare Diagnostics Regional offices on September 2, 2016 both in the United States and outside of the United States, for communication with affected customers e: the bias of the chloride values. US accounts sent hard copy via Federal Express and requested to: Complete and return the Field Correction Effectiveness Check form attached to this letter within 7 days. Review this letter with your Medical Director. Questions:Contact your Siemens Customer Care Center or your local Siemens technical support representative.
Quantity in commerce
689 cartridges
Distribution
Nationwide: Foreign: Canada, Bahamas,Germany, Japan, Mexico.Belgium, Italy,Spain, Korea, France, Austria, Korea, France, Poland, Australia, china, AE, Bangladesh, Columbia, Czechkoslovakia, Bosnia
Product codes and lots
Serial Numbers:2610900214 through 2621001325 Expiry Dates: Sept 13, 2016 through Dec. 6, 2016

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2912-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2912-2016Class IIITerminatedVoluntary: Firm initiated
Z-2913-2016Class IIITerminatedVoluntary: Firm initiated